FDA Adverse Event Death Summary report: N

HEATER-COOLER SYSTEM 3T

MDR report key: 17236603 · Received June 30, 2023

Report

Report Number
9611109-2023-00294
Event Type
Death
Date Received
June 30, 2023
Date of Event
January 10, 2012
Report Date
February 28, 2024
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
Removal / Correction Number
Z-2076/2081-2015
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H10: THROUGH FOLLOW UP COMMUNICATION, LIVANOVA LEARNED THAT POSSIBLE INVOLVED DEVICE SERIAL NUMBERS COULD BE THE FOLLOWING: SNS (B)(6) (MANUFACTURED ON 14 MARCH 2007) OR (B)(6) (MANUFACTURED ON 16 MARCH 2007) OR (B)(6) (MANUFACTURED ON 20 SEPTEMBER 2012) (MODEL: 16-02-80). NO OTHER INFORMATION HAS BEEN MADE AVAILABLE FROM THE CUSTOMER. DHR REVIEW OF POSSIBLE INVOLVED DEVICE SERIAL NUMBERS HAS BEEN COMPLETED AND DID NOT IDENTIFY ANY DEVIATIONS OR NON-CONFORMITIES RELEVANT TO THE REPORTED ISSUE. SOURCE OF PATIENT CONTAMINATION REMAINS UNKNOWN AND THE LIVANOVA DEVICE INVOLVEMENT AS WELL. ANYWAY, LIVANOVA ALREADY IMPLEMENTED A STRATEGY TO DECREASE THE PROBABILITY OF BACTERIA GROW IN THE HC DEVICE BY APPLYING MULTIPLE MEASURES IMPLEMENTED OVER THE PAST FEW YEARS THROUGH DEDICATED CAPA AND FIELD ACTION. NO DEVICE, BETWEEN THE ONES POSSIBLE INVOLVED AND IN USE AT THE HOSPITAL AT THE TIME OF SURGERIES ((B)(6) 2012), WAS UPGRADED WITH VACUUM AND SEALING KIT. THE RISK IS IN THE ACCEPTABLE REGION. NO SPECIFIC ACTION WAS CURRENTLY DEEMED NECESSARY. LIVANOVA WILL KEEP MONITORING THE MARKET.

Additional Manufacturer Narrative · 0

G.5. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). H.9. LIVANOVA DEUTSCHLAND IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. H10: LIVANOVA DEUTSCHLAND MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN DUBLIN, IRELAND. LIVANOVA INITIATED AN INVESTIGATION AND IS FOLLOWING UP WITH THE CUSTOMER. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 0

SEE INITIAL REPORT

Description of Event or Problem · 0

LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A MALE PATIENT UNDERGOING A MITRAL VALVE REPLACEMENT AND CARDIOPULMONARY BYPASS ON (B)(6) 2012 WAS FOUND TO BE INFECTED BY MYCOBACTERIUM CHIMAERA AND DIED ON (B)(6) 2013 FOR AN ENDOCARDITIS. HEATER-COOLER 3T WAS USED IN THIS SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
855319 HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND 16-02-80

Patients

Seq Age Sex Outcome Treatment
1 NA Male Death