FDA Adverse Event
Malfunction
Summary report: N
PENTAX
MDR report key: 17235040
·
Received June 30, 2023
Report
- Report Number
- 9610877-2023-56006
- Event Type
- Malfunction
- Date Received
- June 30, 2023
- Date of Event
- June 21, 2023
- Report Date
- June 30, 2023
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- FAJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL FB-15V IS AVAILABLE IN THE USA WITH A 510K NUMBER K951199. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE OCULAR LENS CLOUDY (NOT CLEAR VIEW). BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE MOISTURE CONDENSATION IN THE OCULAR LENS. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
Description of Event or Problem · 0
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(CLOUDY).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1823592 | PENTAX | FIBEROPTICCYSTOSCOPE 2.1C 4.9TP 400L ROP | FAJ | HOYA CORPORATION PENTAX TOKYO OFFICE | FCY-15RBS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |