MAESTRO TOTAL WRIST RADIAL STEM 7.5MM X 40MM
Report
- Report Number
- 1825034-2010-00214
- Event Type
- Injury
- Date Received
- June 15, 2010
- Date of Event
- August 11, 2009
- Report Date
- March 25, 2010
- Manufacturer
- BIOMET ORTHOPEDICS
- Product Code
- JWJ
- PMA / PMN Number
- K042032
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
THIS EVENT WAS INITIALLY REPORTED ON APRIL 19, 2010 AS A REVISION OF A MAESTRO WRIST ON MEDWATCH REPORT NUMBER 1825034-2010-00122. AT THE TIME, NO PRODUCT IDENTIFICATION WAS AVAILABLE. SUBSEQUENTLY, PRODUCT IDENTIFICATION HAS BEEN PROVIDED AND ADDITIONAL INFORMATION HAS BEEN REPORTED BY THE SURGEON STATING THAT THE REASON FOR REVISION WAS DUE TO INFECTION AND NOT AN IMPLANT ISSUE.THE USER FACILITY IS OUTSIDE OF THE UNITED STATES. NO MEDWATCH REPORT WAS RECEIVED. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR. (B) (4).
CATALOG # - WAS MISSING A DIGIT, CORRECTED TO 180182.
IT WAS REPORTED THAT PATIENT UNDERWENT A WRIST PROCEDURE AND WAS SUBSEQUENTLY REVISED FOR NON-PROSTHETIC RELATED REASONS ON (B) (6) 2009. DURING THE REVISION PROCEDURE, THE SURGEON FOUND EVIDENCE OF METALLOSIS IN THE AREA WHERE THE TITANIUM SCREW AND THE CARPAL PLATE INTERFACE. THE SURGEON REPORTED THE METALLOSIS WAS DUE TO THE FRETTING OF THE SCREW AND THE CARPAL PLATE. IT WAS FURTHER REPORTED BY THE SURGEON THAT THE REVISION WAS DUE TO INFECTION AND NOT DUE TO IMPLANT ISSUES.
IT WAS REPORTED THAT PATIENT UNDERWENT A WRIST PROCEDURE AND WAS SUBSEQUENTLY REVISED FOR NON-PROSTHETIC RELATED REASONS ON (B) (6) 2009. DURING THE REVISION PROCEDURE, THE SURGEON FOUND EVIDENCE OF METALLOSIS IN THE AREA WHERE THE TITANIUM SCREW AND THE CARPAL PLATE INTERFACE. THE SURGEON REPORTED THE METALLOSIS WAS DUE TO THE FRETTING OF THE SCREW AND THE CARPAL PLATE. IT WAS FURTHER REPORTED BY THE SURGEON THAT THE REVISION WAS DUE TO INFECTION AND NOT DUE TO IMPLANT ISSUES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAESTRO TOTAL WRIST RADIAL STEM 7.5MM X 40MM | PROSTHESIS, WRIST | JWJ | BIOMET ORTHOPEDICS | N/A | 294070 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |