FDA Adverse Event
Malfunction
Summary report: N
ORTHO SUMMIT SAMPLE HANDLING SYSTEM
MDR report key: 1722876
·
Received June 15, 2010
Report
- Report Number
- 2250051-2010-00145
- Event Type
- Malfunction
- Date Received
- June 15, 2010
- Date of Event
- May 26, 2010
- Report Date
- June 15, 2010
- Manufacturer
- HAMILTON BONADUZ AG
- Product Code
- JTC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT HAS BEEN INVESTIGATED. THE FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT. THE FSE IDENTIFIED THE PROBLEM WAS DUE TO A STICKY TIP CLAMP AT POSITION 6. THE FSE REPLACED THE TIP CLAMP. FSE PERFORMED SPLLD CHECKING PROCEDURE AND TESTED SUMMIT WITH OAS USER SOFTWARE. THE INSTRUMENT WAS TESTED WITHOUT PROBLEM AND WAS RETURNED TO EXPECTED OPERATION.
Description of Event or Problem · 1
THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT REPORTEDLY DID NOT PIPETTE SAMPLE AND DID NOT GENERATE AN ERROR MESSAGE.NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT.THE REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ORTHO SUMMIT SAMPLE HANDLING SYSTEM | SAMPLE PIPETTER | JTC | HAMILTON BONADUZ AG | 936400 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |