FDA Adverse Event Injury Summary report: N

PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM

MDR report key: 17226345 · Received June 29, 2023

Report

Report Number
2024168-2023-06951
Event Type
Injury
Date Received
June 29, 2023
Date of Event
May 31, 2023
Report Date
August 23, 2023
Manufacturer
ABBOTT VASCULAR
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

VISUAL INSPECTIONS WERE PERFORMED ON THE RETURNED DEVICE. THE REPORTED A NEEDLE TO CUFF MISS WAS CONFIRMED. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. THE REPORTED DIFFICULTIES AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING. D4; LOT # UPDATED FROM 3022141 TO 3022241. H4: MANUFACTURING DATE CORRECTED.

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT SUTURE PLACEMENT IN THE COMMON FEMORAL ARTERY WAS ATTEMPTED WITH A PROSTYLE DEVICE USING THE PRE-CLOSE TECHNIQUE VIA AN 8F SHEATH HOLE PRIOR TO AN TRANSCATHETER AORTIC VALVE IMPLANTATION (TAVI) INTERVENTIONAL PROCEDURE. REPORTEDLY, A CUFF MISS [SUTURE RETRIEVAL ISSUE] OCCURRED. THE SUTURES OF TWO NEW PROSTYLE DEVICES WERE SUCCESSFULLY PRE-PLACED. THE TAVI PROCEDURE WAS COMPLETED. HEMOSTASIS WAS ACHIEVED WITH THE PRE-PLACED PROSTYLE SUTURES. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
254563 PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM SUTURE MEDIATED CLOSURE MGB ABBOTT VASCULAR 3022241

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention