FDA Adverse Event Malfunction Summary report: N

BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC)

MDR report key: 17224953 · Received June 29, 2023

Report

Report Number
2647876-2023-00060
Event Type
Malfunction
Date Received
June 29, 2023
Date of Event
April 11, 2023
Report Date
September 23, 2023
Manufacturer
BD CARIBE, LTD
Product Code
MDB
UDI-DI
00382904420239
PMA / PMN Number
SEE H.10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

G.5. PMA / 510(K)#: K113558, K222591. H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6. INVESTIGATION SUMMARY: CUSTOMER REPORTED A POSITIVE ID RESULT FOR BACTEC MEDIA, WHILE USING BIOFIRE FILMARRAY® BLOOD CULTURE IDENTIFICATION. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH THE BACTEC PRODUCT BASED ON OUR INTERNAL PROCEDURES AND THE INTENDED USE OF THE PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED AND ONLY THE CURRENT COMPLAINT WAS FOUND RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS. AS PER PRODUCT INSERT A GRAM-STAINED SMEAR FROM CULTURE MEDIUM MAY CONTAIN SMALL NUMBERS OF NONVIABLE ORGANISMS DERIVED FROM MEDIA CONSTITUENTS, STAINING REAGENTS, IMMERSION OIL, GLASS SLIDES, AND SPECIMENS USED FOR INOCULATION. DUE TO THE NATURE OF BIOLOGICAL MATERIALS IN MEDIA PRODUCTS AND INHERENT ORGANISM VARIABILITY, THE USER SHOULD BE COGNIZANT OF POTENTIAL VARIABLE RESULTS IN THE RECOVERY OF CERTAIN MICROORGANISMS. MOLECULAR TESTS PERFORMED ON POSITIVE BLOOD CULTURES WILL DETECT BOTH VIABLE AND NON-VIABLE ORGANISMS COMMONLY FOUND IN CULTURE MEDIA. THEREFORE, MOLECULAR TEST RESULTS SHOULD BE EVALUATED IN CONJUNCTION WITH GRAM STAIN RESULTS IN ACCORDANCE WITH STANDARD-OF-CARE PRACTICES AS WELL AS MANUFACTURER¿S INSTRUCTIONS FOR USE. CATALOG: 442023. BATCH NO.: UNKNOWN. CUSTOMER REPORTED A POSITIVE ID RESULT FOR BACTEC MEDIA, WHILE USING BIOFIRE FILMARRAY® BLOOD CULTURE IDENTIFICATION PANELS. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. INVESTIGATION CANNOT BE CONDUCTED TO THE RETENTION SAMPLES SINCE THE LOT NUMBER IS UNKNOWN. A COMPLAINT HISTORY REVIEW CANNOT BE CONDUCTED RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE SINCE BATCH NUMBER IS UNKNOWN. THE BATCH HISTORY RECORD COULD NOT BE REVIEWED AS THE LOT NUMBER IS UNKNOWN NONETHELESS BATCH HISTORY RECORDS ARE ALWAYS REVIEWED PRIOR TO PRODUCT RELEASE. COMPLAINT IS UNCONFIRMED. AS PER PRODUCT INSERT A GRAM-STAINED SMEAR FROM CULTURE MEDIUM MAY CONTAIN SMALL NUMBERS OF NONVIABLE ORGANISMS DERIVED FROM MEDIA CONSTITUENTS, STAINING REAGENTS, IMMERSION OIL, GLASS SLIDES, AND SPECIMENS USED FOR INOCULATION. DUE TO THE NATURE OF BIOLOGICAL MATERIALS IN MEDIA PRODUCTS AND INHERENT ORGANISM VARIABILITY, THE USER SHOULD BE COGNIZANT OF POTENTIAL VARIABLE RESULTS IN THE RECOVERY OF CERTAIN MICROORGANISMS. MOLECULAR TESTS PERFORMED ON POSITIVE BLOOD CULTURES WILL DETECT BOTH VIABLE AND NON-VIABLE ORGANISMS COMMONLY FOUND IN CULTURE MEDIA. THEREFORE, MOLECULAR TEST RESULTS SHOULD BE EVALUATED IN CONJUNCTION WITH GRAM STAIN RESULTS IN ACCORDANCE WITH STANDARD-OF-CARE PRACTICES AS WELL AS MANUFACTURER¿S INSTRUCTIONS FOR USE.

Description of Event or Problem · 0

REPORT 2 OF 17 IT WAS REPORTED THAT WITH USE OF BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) THERE WAS A MOLECULAR FALSE POSITIVE RESULT. NO ERRONEOUS RESULTS REPORTED. THERE WAS NO PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "PT # 2, 95-CB23-138123: LOT #: 2333876, SEQUENCE #: 449257120509, INSTRUMENT S/N: FT (B)(6), NO ERRONEOUS RESULTS REPORTED, BCID RESULT: C. TROPICALIS, STREP SPP., STREP PNEUMONIAE, BCID2, LOT #: 2RBT23, GRAM STAIN: GPC IN PAIRS, CULTURE GREW : STREP PNEUMONIAE".

Description of Event or Problem · 0

REPORT 2 OF 17: IT WAS REPORTED THAT WITH USE OF BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) THERE WAS A MOLECULAR FALSE POSITIVE RESULT. NO ERRONEOUS RESULTS REPORTED. THERE WAS NO PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "PT # 2, 95-CB23-138123: LOT #: 2333876. SEQUENCE #: (B)(4). INSTRUMENT S/N: (B)(6). NO ERRONEOUS RESULTS REPORTED. BCID RESULT: C. TROPICALIS, STREP SPP., STREP PNEUMONIAE. BCID2, LOT #: 2RBT23. GRAM STAIN: GPC IN PAIRS. CULTURE GREW : STREP PNEUMONIAE".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1627632 BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) SYSTEM, BLOOD CULTURING MDB BD CARIBE, LTD 442023 2333876 00382904420239

Patients

Seq Age Sex Outcome Treatment
1 Unknown