FDA Adverse Event Malfunction Summary report: N

CARDIOSAVE HYBRID W/ E/F PLUG

MDR report key: 17221783 · Received June 28, 2023

Report

Report Number
2249723-2023-03008
Event Type
Malfunction
Date Received
June 28, 2023
Date of Event
June 20, 2023
Report Date
April 9, 2024
Manufacturer
DATASCOPE CORP. - MAHWAH
Product Code
DSP
UDI-DI
10607567108414
PMA / PMN Number
K112372
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE FSE THAT ENCOUNTERED THE ISSUE REPLACED THE RIGHT DISPLAY HINGE (0105-00-0138-02) AND THE LEFT DISPLAY HINGE (0105-00-0138-01). THE FSE PERFORMED FUNCTIONAL AND SAFETY CHECKS. THE MAQUET FAILURE ANALYSIS AND TESTING DEPT.(FAT) RECEIVED HINGES PART NUMBER 0105-00-0138-01 AND 0105-00-0138-02 WITH A REPORTED UNIT FAILURE OF THE TOP SCREEN BEING DIFFICULT TO OPEN. THE FAT PERFORMED A VISUAL INSPECTION AND FOUND THE PARTS TO BE IN A WORN CONDITION. THE FAT INSTALLED THE HINGES IN CARDIOSAVE TEST FIXTURE AND TESTED THEM TO FACTORY SPECIFICATIONS PER THE CARDIOSAVE SERVICE MANUAL. IT WAS FOUND THAT THE TOP SCREEN WAS DIFFICULT TO OPEN DUE TO THE HINGES BEING SEIZED UP. FAILURE WAS REPLICATED BUT NO ROOT CAUSE DEFINED. RETAINING THE PART IN THE FAILURE ANALYSIS AND TESTING DEPARTMENT PER PROCEDURE.

Additional Manufacturer Narrative · 0

DUE TO CHARACTER RESTRICTIONS IN BLOCK E1 TELEPHONE NUMBER : (B)(6). A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE (PM) PERFORMED BY A GETINGE SERVICE TERRITORY MANAGER (STM) , THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) UNIT TOP SCREEN WAS DIFFICULT TO OPEN. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1321939 CARDIOSAVE HYBRID W/ E/F PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. - MAHWAH 0998-00-0800-55 10607567108414

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown