ENO
Report
- Report Number
- 1000165971-2023-00485
- Event Type
- Malfunction
- Date Received
- June 28, 2023
- Date of Event
- May 26, 2023
- Report Date
- September 12, 2023
- Manufacturer
- MICROPORT CRM S.R.L.
- Product Code
- LWP
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EI
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H11. CORRECTED DATA: D2. TYPE OF DEVICE CHANGED TO LWP - IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT), SINCE LWS WAS CHOSEN BY MISTAKE AT INITIAL REPORT. G3.3. DATE RECEIVED BY MANUFACTURER CHANGED TO 31/08/2023 AS R&D ANALYSIS RESULTS RECEIVED.
THE DEVICE INVOLVED IN THIS MDR REPORT IS NOT APPROVED IN THE UNITED STATES; HOWEVER, IT IS SIMILAR TO A DEVICE MANUFACTURED BY MICROPORT CRM THAT WAS CLEARED OR APPROVED BY FDA FOR MARKETING IN THE UNITED STATES. PRELIMINARY INVESTIGATION LED TO THE FOLLOWING CONCLUSION: THE VENTRICULAR LEAD (NON MICROPORT CRM PRODUCT) MOST LIKELY IS FRACTURED AND NEEDS TO BE REPLACED. VENTRICULAR OVERSENSING , HIGH VENTRICULAR IMPEDANCE , VENTRICULAR BURSTS SHOWING ARTEFACTS CHARACTERISTIC FOR LEAD FRACTURE, IN ADDITION HIGH FREQUENCY NOISE OVERSENSING WAS OBSERVED THAT MIGHT BE RELATED TO THE SUSPECTED LEAD FRACTURE BUT NEEDS FURTHER INVESTIGATION.
REPORTEDLY, NOISE OVERSENSING WAS DETECTED ON THE VENTRICULAR LEAD DURING DEVICE FOLLOW UP. FURTHER INVESTIGATION AND ARM MANIPULATION REPRODUCED FURTHER VENTRICULAR OVERSENSING.
REPORTEDLY, NOISE OVERSENSING WAS DETECTED ON THE VENTRICULAR LEAD DURING DEVICE FOLLOW UP. FURTHER INVESTIGATION AND ARM MANIPULATION REPRODUCED FURTHER VENTRICULAR OVERSENSING
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1231666 | ENO | IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) | LWP | MICROPORT CRM S.R.L. | ENO SR | ||
| 1782919 | ENO | IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) | LWP | MICROPORT CRM S.R.L. | ENO SR |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |