CAVITRON 300 SERIES G310
Report
- Report Number
- 2424472-2023-00034
- Event Type
- Malfunction
- Date Received
- June 28, 2023
- Report Date
- June 28, 2023
- Manufacturer
- DENTSPLY LLC
- Product Code
- ELC
- PMA / PMN Number
- K150535
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE LACK OF WATER FLOW HAS CAUSED AN OVERHEATING INSERT. SINCE AN OVERHEATING INSERT COULD NECESSITATE MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION, THIS MALFUNCTION WOULD BE LIKELY TO CAUSE/CONTRIBUTE TO A SERIOUS INJURY SHOULD IT RECUR. AS SUCH, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. INVESTIGATION RESULTS: NOTES: 06/27/2023, REPAIR TECH: JCB. W4O WATER SOL,ISO VALVE LEAKING NO WATER FLOW DUE TO A CLOGGED WATER SOLENOID, DEBRIS BUILDUP IN THE WATER SUPPLY HOSE, DEBRIS BUILDUP IN THE WATER FILTER. REPLACED DAMAGED/WORN COMPONENTS AND RECALIBRATED UNIT TO FACTORY SPECS. ADDED NEW STACK BOARD.
IN THIS EVENT IT IS REPORTED THAT CAVITRON 300 SERIES G310 OVERHEATED DURING USE, NO INJURY. INVESTIGATION FOUND THAT THERE IS NO WATER FLOW FOR THIS DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 158347 | CAVITRON 300 SERIES G310 | SCALER, ULTRASONIC | ELC | DENTSPLY LLC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |