FDA Adverse Event Malfunction Summary report: Y

EUROMI REUSABLE CANNULA

MDR report key: 17207114 · Received June 27, 2023

Report

Report Number
3007697975-2018-00010
Event Type
Malfunction
Date Received
June 27, 2023
Date of Event
October 10, 2018
Report Date
June 27, 2023
Manufacturer
EUROMI SA
Product Code
GEA
PMA / PMN Number
878.4800
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE CANNULAS HAVE NO LINEARITY OR CORROSION DEFECTS. THEY WERE USED ONLY ONCE IN A DEMONSTRATION. THE BATCH OF TUBING USED TO MAKE THE CANNULA IS COMPLIANT. A FOLD OF THIS TYPE AT THE CANNULA OCCURS: AFTER AN EXCESSIVE LOAD ON THE CANNULA (STRESS AMPLITUDE), OR IF THE CANNULA TUBE IS TOO THIN, OR IF THE TENSILE STRENGTH RM IS TOO LOW (< 460 MPA), OR IF THE DESIGN OF THE CANNULA IS NOT SUITABLE: THE SURFACE OF THE HOLES IS TOO LARGE, THUS CREATING A FRAGILITY OF THE CANNULA. THE OGIVE (TIP) IS CRUSHED AT THE HOLES, AS WELL AS THE TUBE. DESIGN REVIEW TO REINFORCE THE MECHANICAL RESISTANCE OF THE TUBE (DECREASE IN WIDTH AND HEIGHT). A BATCH RECALL IS NOT NECESSARY BECAUSE ALTHOUGH THE TUBE FOLDS AT THE HOLES WHEN INSERTED INTO THE PATIENT'S BODY, THE CANNULA DOES NOT BREAK. IN ADDITION, THE CANNULA IS CHECKED BEFORE EACH PROCEDURE (CF. INSTRUCTIONS). THE MERCEDES MODEL HAS BEEN AVAILABLE SINCE EUROMI OBTAINED THE CE MARKING FOR LIPOSUCTION CANNULAS (2009). CLAIMS AND CASES OF VIGILANCE ARE LIMITED WHEN CONSIDERING THE PERIOD 2011 - 2018 (PMS DATA AVAILABLE). THE DELAY IN MDR SUBMISSION IS DUE TO EMPLOYEE TURNOVER IN THE QUALITY DEPARTMENT. INITIALLY, REPORT WAS SENT IN VIA THE ESG PORTAL IN SEPTEMBER 2022 - DUE TO THE WRONG FORMAT BEING USED, IT IS RESENT.

Description of Event or Problem · 0

THE MERCEDES CANNULA FOLDED AT THE HOLES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1478069 EUROMI REUSABLE CANNULA CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY GEA EUROMI SA MERCEDES TUNGSTEN 3*300 18AHAC309

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other