Narratives
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION.
DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF DEFLATION WAS RECEIVED ON JULY 28, 2023, WITH LOT NUMBER 2010172. BASED ON THE DEVICE ANALYSIS GRID, THE ASSESSMENTS OF THE COMPLAINT ARE: ¿ DEFLATION: OBSERVED AN OPENING ON ANTERIOR ASSESSED AS AN UNIDENTIFIED (TEAR) OPENING (SHELL THICKNESS WITHIN SPEC) AND OBSERVED A BROKEN-ON PLUG STRAP ASSESSED AS AN UNIDENTIFIED (TEAR) OPENING. ADDITIONAL OBSERVATIONS: ¿ WHITE PARTICLES OBSERVED INSIDE THE DEVICE. NO FURTHER ACTIONS ARE REQUIRED AS THE DEVICE WAS IMPLANTED.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED AND REPLACED.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 833640 | STYLE 68 SALINE FILLED BREAST IMPLANT | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | ALLERGAN (COSTA RICA) | 2010172 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR | Female | Required Intervention |