FDA Adverse Event Malfunction Summary report: N

HALOGEN LIGHT SOURCE

MDR report key: 17201167 · Received June 26, 2023

Report

Report Number
3002808148-2023-06380
Event Type
Malfunction
Date Received
June 26, 2023
Date of Event
May 29, 2023
Report Date
August 11, 2023
Manufacturer
SHIRAKAWA OLYMPUS CO., LTD.
Product Code
FAJ
UDI-DI
04953170024528
PMA / PMN Number
K993041
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION, AND THE CUSTOMER'S ALLEGATION WAS CONFIRMED; DUE TO A FAULTY TRANSFORMER UNIT. THE ADDITIONAL EVALUATION FINDINGS ARE AS FOLLOWS: THE WASHER UNDER THE SCREW ON REAR PANEL WAS MISSING, AND THE LAMP HARNESS WAS DEFORMED. THE INVESTIGATION IS ONGOING, AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION OR IF ANY ADDITIONAL INFORMATION IS PROVIDED BY THE USER FACILITY.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THAT THE EVENT OCCURRED DUE TO A FAULTY TRANSFORMER UNIT. HOWEVER, THE DEFINITIVE ROOT CAUSE WAS UNABLE TO BE DETERMINED. THE LAMP WAS FOUND TO BE A THIRD PARTY LAMP. OLYMPUS ALWAYS RECOMMENDS TO USE OLYMPUS MD-151 HALOGEN LAMP. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

THE OLYMPUS FIELD SERVICE ENGINEER REPORTED (ON BEHALF OF THE CUSTOMER) THAT THE HALOGEN LIGHT SOURCE LAMP DOES NOT LIGHT. THE ISSUE WAS FOUND DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT HARM AND NO PROCEDURAL INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1453861 HALOGEN LIGHT SOURCE HALOGEN LIGHT SOURCE FAJ SHIRAKAWA OLYMPUS CO., LTD. CLK-4 04953170024528

Patients

Seq Age Sex Outcome Treatment
1 Unknown