FDA Adverse Event Injury Summary report: N

SITE MICROSURGICAL SYSTEMS, INC.

MDR report key: 17201 · Received August 31, 1994

Report

Report Number
17201
Event Type
Injury
Date Received
August 31, 1994
Date of Event
August 22, 1994
Report Date
August 31, 1994
Manufacturer
SITE MICROSURGICAL SYSTEMS, INC.
Product Code
HQC
Adverse Event
Yes
Product Problem
Yes
Report Source
User Facility report
Reporter Location
TX, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

PT WAS THE SECOND CASE FOR PHACO CATARACT EXTRACTION USING THE IRRIGATION AND ASPIRATION MODULE. THE FIRST CASE WAS COMPLETED WITHOUT INCIDENT. DURING THE PHACOEMULSIFICATION PHASE, AFTER THREE PARTS OF THE FRACTURE WERE MADE, AND WHILE BEGINNING ON THE FOURTH ARM OF THE FRACTURE, SOME INSTANTANEOUS CORNEAL WHITENING WAS NOTED. PHACOEMULSIFICATION WAS IMMEDIATELY STOPPED AND THE PHACO TIP REMOVED FROM THE EYE. THE LINES WERE CHECKED AND IT APPEARED THAT IRRIGATION WAS NORMAL THROUGH THE INSTRUMENT. HAVING NO OTHER RECOURSE AT THAT POINT, SINCE THE PHACOEMULSIFICATION HAD ALREADY BEGAN AND THERE WAS A SMALL PUPIL, THE TIP WAS PLACED BACK IN THE EYE AND THE REMAINDER OF THE NUCLEUS WAS PHACOEMULSIFIED WITHOUT COMPLICATION. A SCLERAL PATCH WAS PERFORMED ON THE PT. THE PHYSICIAN SUSPECTED THAT A REDUCED IRRIGATION FLOW THROUGH THE HANDPIECE WAS THE CAUSE OF THE OVERHEATING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SITE MICROSURGICAL SYSTEMS, INC. PHACO MODULE HQC SITE MICROSURGICAL SYSTEMS, INC. 501-3760

Patients

Seq Age Sex Outcome Treatment
1 85 YR Required Intervention