60" CENTURY TRANSFER CART
Report
- Report Number
- 3005899764-2010-00012
- Date Received
- June 11, 2010
- Date of Event
- May 13, 2010
- Report Date
- June 11, 2010
- Manufacturer
- STERIS MEXICO, S. DE R.L. DE C.V.
- Product Code
- FLE
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- OTHER
Narratives
THE USER FACILITY DECLINED STERIS' OFFER OF IN-SERVICE TRAINING.
A STERIS SERVICE TECHNICIAN WENT ON-SITE TO INSPECT THE TRANSFER CART, HOWEVER WHEN HE ARRIVED, HE FOUND THAT THE FACILITY PERSONNEL HAD ALREADY RE-INSERTED THE WHEEL INTO THE CART LEG. UPON INSPECTION OF THE CART, THE TECHNICIAN FOUND THAT THE WHEEL HAD NOT BEEN PROPERLY RE-INSERTED. THE TECHNICIAN RE-INSERTED THE WHEEL CORRECTLY, TIGHTENED THE CASTER AND VERIFIED THAT ALL OTHER CASTERS WERE PROPERLY SECURED. THE CART WAS RETURNED TO SERVICE AND NO FURTHER ISSUES HAVE BEEN REPORTED WITH THIS EQUIPMENT. THE STERIS SERVICE TECHNICIAN HAS OBSERVED IN THE PAST THE USER FACILITY PERSONNEL INCORRECTLY REMOVING THE TRANSFER CART FROM THE STERILIZER. SPECIFICALLY, EMPLOYEES WERE INSTRUCTED TO UNLOAD THE CART BY PULLING THE CART STRAIGHT OUT OF THE STERILIZER AND ARE NOT TO PULL OR PUSH ON THE SIDES OF THE CART TO REMOVE. STERIS HAS DETERMINED THE INCIDENT OCCURRED DUE TO USER ERROR AND HAS REINFORCED THE IMPORTANCE OF PULLING THE CART STRAIGHT OUT OF THE STERILIZER AND RECOMMENDED A VISUAL CHECK OF ALIGNMENT PRIOR TO PLACING OR REMOVING A CART FROM THE STERILIZER. STERIS HAS OFFERED THE FACILITY FORMAL REFRESHER IN-SERVICE TRAINING ADDRESSING THE PROPER LOADING AND UNLOADING PROCESS FOR TRANSFER CARTS AND ARE AWAITING THE USER FACILITY'S RESPONSE.
THE USER FACILITY REPORTED THAT WHEN AN EMPLOYEE WAS LOADING A STERILIZER THE RIGHT REAR WHEEL OF THE TRANSFER CART CAME OFF, CAUSING THE CART RACK TO FALL TO THE FLOOR WHILE THE EMPLOYEE WAS HOLDING THE HANDLE OF THE LOADING CART RACK. THE EMPLOYEE WENT TO THE FACILITY EMERGENCY ROOM AND WAS GIVEN MEDICATION FOR PAIN. THE FACILITY REPORTED THE EMPLOYEE MISSED NO TIME FROM WORK AND SUSTAINED NO INJURIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 60" CENTURY TRANSFER CART | TRANSFER CARRIAGE | FLE | STERIS MEXICO, S. DE R.L. DE C.V. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |