FDA Adverse Event Malfunction Summary report: N

CAVITRON JET PLUS W/TAPON -115V G137

MDR report key: 17198066 · Received June 26, 2023

Report

Report Number
2424472-2023-00033
Event Type
Malfunction
Date Received
June 26, 2023
Report Date
July 31, 2023
Manufacturer
DENTSPLY LLC
Product Code
ELC
PMA / PMN Number
K052334
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE LACK OF WATER FLOW HAS CAUSED AN OVERHEATING INSERT. SINCE AN OVERHEATING INSERT COULD NECESSITATE MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION, THIS MALFUNCTION WOULD BE LIKELY TO CAUSE/CONTRIBUTE TO A SERIOUS INJURY SHOULD IT RECUR. AS SUCH, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH HAS NOT BEEN RETURNED AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.

Additional Manufacturer Narrative · 0

SAK EMH FOUND HPC WATER FLOW NON FUNCTIONAL, NO WATER FLOW. REPLACED HPC WITH NEW, ALSO INSTALLED NEW WATER FILTER AND PINCH TUBE, UNIT THEN TUNED/CALIBRATED AND ALL FUNCTIONS TESTED QA-RB. NO JET-MATE WITH UNIT.

Description of Event or Problem · 0

WHILE USING A CAVITRON JET PLUS G137, THEY ALLEGE THAT THEY HAD LOW WATER PRESSURE AND THE HANDPIECE WAS GETTING HOT. NO INJURY RESULTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
841818 CAVITRON JET PLUS W/TAPON -115V G137 SCALER, ULTRASONIC ELC DENTSPLY LLC

Patients

Seq Age Sex Outcome Treatment
1 Unknown