FDA Adverse Event Malfunction Summary report: N

RESPIRATORY HUMIDIFIER

MDR report key: 1719568 · Received May 31, 2010

Report

Report Number
9611451-2010-00342
Event Type
Malfunction
Date Received
May 31, 2010
Report Date
May 3, 2010
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD.
Product Code
BTT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE HC550 RESPIRATORY HUMIDIFIER IS NOT SOLD IN THE UNITED STATES, BUT IS SIMILAR TO A PRODUCT THAT IS SOLD IN THE UNITED STATES. THE 510 (K) NUMBER OF THE SIMILAR PRODUCT IS K033710. METHOD: THE COMPLAINT HC550 RESPIRATORY HUMIDIFIER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE'S OFFICE IN (B)(4) FOR TESTING. THE HUMIDIFIER WAS TESTED IN ACCORDANCE WITH THE PRODUCT TECHNICAL MANUAL. RESULTS: THE HC550 HUMIDIFIER FUNCTIONED PROPERLY DURING TESTING AND NO FAULT WAS FOUND WITH THE DEVICE. CONCLUSION: NO FAULT WAS FOUND WITH THE DEVICE AND THE REPORTED FAULT COULD NOT BE REPLICATED. CONDENSATE IN THE HUMIDIFICATION SYSTEM, ALTHOUGH NOT PREFERRED, IS AN EXPECTED SIDE EFFECT OF HEATED PASS-OVER HUMIDIFICATION SYSTEMS IN MANY CONDITIONS, AND MAY VARY BETWEEN LIGHT MISTING TO WATER DROPLETS THAT FORM ON THE WALL OF COOL BREATHING CIRCUIT TUBING. THE AMOUNT OF CONDENSATE IN THE VENTILATION SYSTEM IS INFLUENCED BY MULTIPLE SET-UP AND ENVIRONMENTAL FACTORS. ATTEMPTS HAVE BEEN MADE TO OBTAIN FURTHER INFORMATION FROM THE CUSTOMER INCLUDING THE AMOUNT OF CONDENSATION OBSERVED IN THE SYSTEM TO ASSIST US WITH THE ANALYSIS AND IDENTIFICATION OF A POSSIBLE ROOT CAUSE OF THE EVENT AS REPORTED; HOWEVER, NO FURTHER INFORMATION HAS BEEN FORTHCOMING. WITHOUT FURTHER INFORMATION SURROUNDING THE ISSUE, WE DO NOT KNOW THE EXTENT OF THE CONDENSATION ISSUE EXPERIENCED AND EXACTLY WHAT CAUSED THE REPORTED INCIDENT. IT IS HOWEVER POSSIBLE THAT SET-UP AND ENVIRONMENTAL CONDITIONS MAY HAVE CONTRIBUTED TO THE CAUSE OF THE ISSUE.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT THERE WAS TOO MUCH CONDENSATION IN A BREATHING CIRCUIT. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RESPIRATORY HUMIDIFIER BTT FISHER & PAYKEL HEALTHCARE LTD. HC550 090119

Patients

Seq Age Sex Outcome Treatment
1