FS OPT B-KET-NO CODING
Report
- Report Number
- 2954323-2023-26834
- Event Type
- Malfunction
- Date Received
- June 23, 2023
- Date of Event
- June 2, 2023
- Report Date
- August 7, 2023
- Manufacturer
- ABBOTT DIABETES CARE INC
- Product Code
- QLG
- PMA / PMN Number
- K193371
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- 003
Narratives
THE PRODUCT HAS BEEN REQUESTED BACK FOR INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
REPEATED ATTEMPTS BY ADC TO RETRIEVE THE PRODUCT WERE UNSUCCESSFUL AND/OR THE CUSTOMER DISCARDED THE PRODUCT. NO PRODUCT HAS BEEN RETURNED. AN EXTENDED INVESTIGATION HAS BEEN PERFORMED FOR THE REPORTED COMPLAINT AND IT HAS BEEN DETERMINED THAT THERE WAS NO INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATIONS. THE DHRS (DEVICE HISTORY REVIEW) FOR THE LIBRE READER WAS REVIEWED AND THE DHRS SHOWED THE LIBRE READER PASSED ALL TESTS PRIOR TO RELEASE. THE DHRS (DEVICE HISTORY REVIEW) FOR THE PRECISION STRIPS WAS REVIEWED AND THE DHRS SHOWED THE PRECISION STRIPS PASSED ALL TESTS PRIOR TO RELEASE. RETAIN TESTING WAS PERFORMED FOR PRECISION STRIPS, AND ALL UNITS PERFORMED IN SPECIFICATION AND PASSED. IF THE PRODUCT IS RETURNED, THE CASE WILL BE RE-OPENED, AND A PHYSICAL INVESTIGATION WILL BE PERFORMED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1944716 | FS OPT B-KET-NO CODING | FLASH GLUCOSE MONITORING SYSTEM | QLG | ABBOTT DIABETES CARE INC | 71953-01 | 45001 86774 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |