FDA Adverse Event Malfunction Summary report: N

PFNA-II BLADE L75 TAN

MDR report key: 17190881 · Received June 23, 2023

Report

Report Number
8030965-2023-07960
Event Type
Malfunction
Date Received
June 23, 2023
Date of Event
May 30, 2023
Manufacturer
SYNTHES GMBH
Product Code
HSB
UDI-DI
07611819394193
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10 ADDITIONAL NARRATIVE: D9: COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. E3: REPORTER IS A J&J SALES REPRESENTATIVE. G4: DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES, BUT IS SIMILAR TO DEVICE MARKETED IN THE USA. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE HISTORY (LOT) PART# 04.027.050S LOT # L820711 MANUFACTURING SITE: WERK BETTLACH RELEASE TO WAREHOUSE DATE: 09 MARCH 2018 SUPPLIER: N/A EXPIRATION DATE: N/A A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED ARTICLE LOT AND NO NON-CONFORMANCES WERE IDENTIFIED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON MAY 30, 2023, THE PATIENT UNDERWENT SURGERY WITH PFNA-II BLADES. WHEN USING A 75MM BLADE, THE BLADE COULD NOT BE LOCKED, SO IT WAS REPLACED WITH ANOTHER BLADE. THE EVENT PRODUCT WAS DISCARDED BY THE NURSE. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) PFNA-II BLADE L75 TAN. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1866483 PFNA-II BLADE L75 TAN ROD, FIXATION, INTRAMEDULLARY HSB SYNTHES GMBH L820711 07611819394193

Patients

Seq Age Sex Outcome Treatment
1 Unknown