INFUSOMAT SPACE
Report
- Report Number
- 9610825-2023-00288
- Event Type
- Malfunction
- Date Received
- June 23, 2023
- Date of Event
- February 8, 2023
- Report Date
- September 20, 2023
- Manufacturer
- B.BRAUN MELSUNGEN AG
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PO
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MELSUNGEN AG INTERNAL REPORT (B)(4). 1. GENERAL INFORMATION: COMPLAINT: (B)(4). ---------------------------------------------------------------- 2. INFORMATION TO THE SAMPLE: 2.1 MODEL: INFUSOMAT SPACE 2.2 ARTICLE NUMBER: 8713050 2.3 SERIAL NUMBER/BATCH: (B)(6) 2.4 SOFTWARE VERSION: N030005 2.5 HOURS OF OPERATION: 13233 2.6 FURTHER INFORMATION: N/A ---------------------------------------------------------------- 3. INVESTIGATION RESULTS: 3.1 HISTORY INSPECTION: THE DEVICE HISTORY FILES WERE READ OUT AND ANALYZED. NO ABNORMALITIES WERE FOUND IN THE DEVICE AND ALARM HISTORY. 3.2 VISUAL INSPECTION: A VISUAL INSPECTION WAS PERFORMED. THE COVER CAPS ON THE SCREW PILLARS, AND THE TECHNICIAN SEAL (351-02-003) ON THE LOWER HOUSING WERE INTACT AND UNDAMAGED. THE DEVICE IS IN A CLEAN STATE AND NO VISIBLE DAMAGE ARE TO LOCATE. 3.3 FUNCTIONAL INSPECTION: A FUNCTIONAL TEST WAS PERFORMED. THE DEVICE PASSED THE SELF-TEST. A SPACE LINE WAS INSERTED, THE PUMP IDENTIFIED THE LINE, AND IT COULD BE SELECTED FROM THE MENU. IT WAS POSSIBLE TO PUT THE PUMP IN OPERATION. 3.4 PRESSURE INSPECTION: IN CHECKING THE DOWNSTREAM SENSOR, THE ELECTRONIC PRESSURE CUT-OFF AND THE MECHANICAL PRESSURE LIMITATION OF THE DEVICE WERE TESTED, ACCORDING TO THE REQUIREMENTS OF THE TECHNICAL SAFETY CHECK. THE DEVICE MATCHES THE REQUIRED VALUES AND STANDARDS. ALL MEASURED VALUES ARE WITHIN OUR SPECIFICATION. 3.5 FLOW RATE INSPECTION: A DELIVERY ACCURACY MEASUREMENT ACCORDING TO IEC 60601-2-24 WAS ARRANGED. HERE A NOMINAL FLOW RATE OF 100 ML/H WAS CHOSEN. THE ASSESSED AVERAGE DEVIATION "A" OF THE SECOND OPERATING HOUR WAS MEASURED AND RESULTED IN A VALUE OF -0,76%. ((ACCURACY OF SET DELIVERY RATE SHOULD BE: ± 5 % ACCORDING TO IEC/EN 60601-2-24) THE DEVICE MATCHES THE REQUIRED VALUES AND STANDARDS. ALL MEASURED VALUES ARE WITHIN OUR SPECIFICATION. 3.6 DISASSEMBLING: DURING THE INVESTIGATION NO FAULTS COULD BE DETECTED, TO INVESTIGATE THE INSIDE OF THE DEVICE, ONLY THE UPPER HOUSING WAS REMOVED. NO DAMAGE OR SOILING COULD BE FOUND. 4. JUDGMENT: 4.1 THE COMPLAINT COULD NOT BE CONFIRMED. SUMMING UP ALL TESTS, THE INFUSOMAT SPACE OPERATES WITHIN OUR SPECIFICATION. NO PRODUCT DEVIATION. NOTE: THIS REPORT IS BEING FILED FOR AN ITEM NUMBER THAT IS NOT SOLD IN THE UNITED STATES, HOWEVER SIMILAR ITEMS ARE SOLD IN THE UNITED STATES BY B. BRAUN MEDICAL, INC.
THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MELSUNGEN AG INTERNAL REPORT 400606449. THE COMPLAINT IS UNDER EVALUATION. A FOLLOW-UP REPORT WILL BE PROVIDED AFTER THE EXAMINATION RESULTS ARE AVAILABLE. NOTE: THIS REPORT IS BEING FILED FOR AN ITEM NUMBER THAT IS NOT SOLD IN THE UNITED STATES, HOWEVER SIMILAR ITEMS ARE SOLD IN THE UNITED STATES BY B. BRAUN MEDICAL, INC.
AS REPORTED BY THE USER FACILITY INFORMATION BY BBM SALES ORGANIZATION IN PORTUGAL: "OVERINFUSION". ACCORDING TO THE CUSTOMER: "IT WAS REPORTED BY THE CUSTOMER TWO INCIDENTS WITH 2 INFUSOMAT EQUIPMENTS, WHERE IT WERE REPORTED DEVIATIONS IN THE ADMINISTRATION HIGHER THAN 30%".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1001539 | INFUSOMAT SPACE | PUMP, INFUSION, | FRN | B.BRAUN MELSUNGEN AG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |