HEATER COOLER 3T
Report
- Report Number
- 9611109-2023-00282
- Event Type
- Death
- Date Received
- June 22, 2023
- Date of Event
- January 14, 2015
- Report Date
- February 29, 2024
- Manufacturer
- LIVANOVA DEUTSCHLAND GMBH
- Product Code
- DWC
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EI
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
D.4. SERIAL NUMBER AND IS UNKNOWN. AS THE SERIAL NUMBER IS UNKNOWN, THE UDI COULD NOT BE DETERMINED. THESE INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT IF MADE AVAILABLE. G.5. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). H.4. AS THE SERIAL NUMBER IS UNKNOWN, THE DEVICE MANUFACTURE DATE COULD NOT BE DETERMINED. THIS INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT IF MADE AVAILABLE. H10: LIVANOVA DEUTSCHLAND MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN IRELAND. LIVANOVA INITIATED AN INVESTIGATION. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
H10: THROUGH FOLLOW UP COMMUNICATION, LIVANOVA LEARNED THAT ALLEGED MALFUNCTION CLAIMED BY THE COMPLAINANT WAS A MICROBIAL INFECTION DISCOVERED IN (B)(6) 2017. DUE TO GENTAMICIN TREATMENT, PATIENT DEVELOPED TOXICITY AFFECTING HIS BALANCE. IN ADDITION, IT WAS LEARNED THAT POSSIBLE INVOLVED DEVICE SERIAL NUMBERS COULD BE THE FOLLOWING: (B)(6) (MODEL: 16-02-80). NO OTHER INFORMATION HAS BEEN MADE AVAILABLE FROM THE CUSTOMER. DEVICE HISTORY RECORDS (DHRS) REVIEW OF POSSIBLE INVOLVED DEVICE SERIAL NUMBERS HAS BEEN COMPLETED AND DID NOT IDENTIFY ANY DEVIATION OR NON-CONFORMITY RELEVANT TO THE REPORTED ISSUE. NO DEVICE, BETWEEN THE ONES POSSIBLE INVOLVED AND IN USE AT THE HOSPITAL AT THE TIME OF SURGERY (2015), WAS UPGRADED WITH VACUUM AND SEALING KIT. NO FURTHER INVESTIGATION IS POSSIBLE. SOURCE OF PATIENT CONTAMINATION REMAINS UNKNOWN AND A DIRECT RELATIONSHIP BETWEEN THE REPORTED ADVERSE EVENT AND THE DEVICE COULD NOT BE ESTABLISHED. THE EVENT WAS IDENTIFIED THROUGH A RETROSPECTIVE REVIEW OF PRODUCT LIABILITY LAWSUITS. THROUGH THIS REVIEW, WE HAVE IDENTIFIED A FEW EVENTS WHERE THE PLAINTIFF HAD FILED A SUIT AGAINST THE COMPANY, AND IT WAS HANDLED BY OUR LEGAL DEPARTMENT, BUT THE UNDERLYING PRODUCT INFORMATION HAD NOT BEEN FORWARDED TO THE COMPLAINT HANDLING UNIT. LIVANOVA OPENED A CAPA TO IDENTIFY ANY POSSIBLE LAWSUITS THAT MIGHT REQUIRE COMPLAINT REPORTING AND TO RETROSPECTIVELY SUBMIT THOSE EVENTS THAT WERE NOT PREVIOUSLY SUBMITTED AND TO PREVENT FUTURE SIMILAR ISSUES FROM OCCURRING.
LIVANOVA DEUTSCHLAND HAS RECEIVED A REPORT OF A DEFECT IN HEATER-COOLER UNIT USED DURING HEART SURGERY.
SEE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1217386 | HEATER COOLER 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND GMBH | 16-02-80 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |