HAHN TAPERED IMPLANT Ø3.5 X 11.5 MM
Report
- Report Number
- 3011649314-2023-00370
- Event Type
- Injury
- Date Received
- June 21, 2023
- Date of Event
- January 31, 2023
- Report Date
- May 6, 2025
- Manufacturer
- PRISMATIK DENTALCRAFT, INC.
- Product Code
- DZE
- PMA / PMN Number
- K143353
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
THE DEVICE HAS NOT BEEN RETURNED. IF/ WHEN THE DEVICE IS RETURNED AN INVESTIGATION WILL BE CARRIED OUT AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS COMPLAINT WILL BE KEPT ON RECORD FOR TRACK AND TRENDING PURPOSES.
THE DEVICE HAS NOT BEEN RETURNED. HOWEVER, THE NON-VISUAL DEVICE EVALUATION HAS BEEN COMPLETED AND THE RESULTS ARE AS FOLLOWS: DHR RESULTS THE DHR WAS REVIEWED FOR HAHN TAPERED IMPLANT LOT# 6113527 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS A REVIEW OF STOCK PRODUCT WAS PERFORMED FOR HAHN TAPERED IMPLANT LOT# 6113527 AND FOUND NO ADDITIONAL PRODUCT IN STOCK. INVESTIGATION METHODS/RESULTS CUSTOMER DID NOT RETURN THE REPORTED DEVICE FOR REVIEW TO DATE. HOWEVER, THE NON-VISUAL DEVICE INVESTIGATION HAS BEEN COMPLETED. ROOT CAUSE "FAILURE TO OSSEOINTEGRATE" IS A COMMON COMPLAINT IN REGARD TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR FAILURE TO OSSEOINTEGRATE IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT, OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. PER THE REPORTED INFORMATION, THE PATIENT HAS A HISTORY OF PERIODONTITIS AND BRUXISM. ADDITIONALLY, THE PATIENT HAS ORAL HYGIENE LISTED AS FAIR. IFU 570 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE CONTRAINDICATIONS SECTION: "HAHN TAPERED IMPLANTS SHOULD NOT BE PLACED IN PATIENTS DISCOVERED TO BE MEDICALLY UNFIT FOR THE INTENDED TREATMENT. PRIOR TO CLINICAL INTERVENTION, PROSPECTIVE PATIENTS MUST BE THOROUGHLY EVALUATED FOR ALL KNOWN RISK FACTORS AND CONDITIONS RELATED TO ORAL SURGICAL PROCEDURES AND SUBSEQUENT HEALING. CONTRAINDICATIONS INCLUDE BUT ARE NOT LIMITED TO: VASCULAR CONDITIONS, UNCONTROLLED DIABETES, CLOTTING DISORDERS, ANTICOAGULANT THERAPY, METABOLIC BONE DISEASE, CHEMOTHERAPY OR RADIATION THERAPY, CHRONIC PERIODONTAL INFLAMMATION, INSUFFICIENT SOFT TISSUE COVERAGE, METABOLIC OR SYSTEMIC DISORDERS ASSOCIATED WITH WOUND AND/OR BONE HEALING, USE OF PHARMACEUTICALS THAT INHIBIT OR ALTER NATURAL BONE REMODELING, ANY DISORDERS WHICH INHIBIT A PATIENT'S ABILITY TO MAINTAIN ADEQUATE DAILY ORAL HYGIENE, UNCONTROLLED PARAFUNCTIONAL HABITS, INSUFFICIENT HEIGHT AND/OR WIDTH OF BONE, AND INSUFFICIENT INTERARCH SPACE." IFU 570 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN PRECAUTION SECTION SURGICAL PROCEDURES: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. FOR BEST RESULTS, PLEASE OBSERVE THE FOLLOWING PRECAUTIONS: ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL, COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO." IFU 570 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." IFU 570 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN."
ADDITIONAL INFORMATION: H6 (TYPE OF INVESTIGATION). CORRECTED INFORMATION: B5, G1, H4, H6 (INVESTIGATION FINDINGS, INVESTIGATION CONCLUSION). CAPA CA-00016. MANUFACTURER REFERENCE: (B)(4). RELATED MW: 3011649314-2023-00369.
ADDITIONAL INFORMATION: A4, A6, B5. CAPA CA-00016. MANUFACTURER REFERENCE: (B)(4). RELATED MW: 3011649314-2023-00369.
IT WAS REPORTED THAT THE HAHN TAPERED IMPLANT FAILED. THE PATIENT'S BONE TYPE IS II, AND THEIR ORAL HYGIENE IS LISTED AS FAIR. THE PATIENT HAS A HISTORY OF PERIODONTITIS AND BRUXISM. THE PATIENT PRESENTED ON (B)(6) 2022 FOR A PRIMARY PROCEDURE ON TOOTH #7. THE PATIENT RETURNED ON (B)(6) 2023 BEFORE THE SECOND STAGE OF SURGERY WITH COMPLAINTS OF PAIN. THE PATIENT LATER RETURNED ON (B)(6) 2023, AND IT WAS DETERMINED THAT THE IMPLANT FAILED TO INTEGRATE, AND THE DEVICE WAS REMOVED.
IT WAS REPORTED THAT THE HAHN TAPERED IMPLANT FAILED. THE PATIENT'S BONE TYPE IS II, AND THEIR ORAL HYGIENE IS LISTED AS FAIR. THE PATIENT PRESENTED ON (B)(6) 2022 FOR A PRIMARY PROCEDURE ON TOOTH #6 AND #7. THE SPECIFIC TOOTH FOR THIS REPORTED IMPLANT LOT WAS NOT INDICATED. THE PATIENT RETURNED ON (B)(6) 2023 BEFORE THE SECOND STAGE OF SURGERY WITH COMPLAINTS OF PAIN. THE PATIENT LATER RETURNED ON (B)(6) 2023, AND IT WAS DETERMINED THAT THE IMPLANT FAILED TO INTEGRATE, AND THE DEVICE WAS REMOVED AND NOT REPLACED. THE SYMPTOMS RESOLVED AFTER DEVICE REMOVAL AND THERE WAS NO PERMANENT PATIENT INJURY. NO ADDITIONAL MEDICAL/SURGICAL PROCEDURE WAS REQUIRED.
IT WAS REPORTED THAT THE HAHN TAPERED IMPLANT FAILED. THE PATIENT'S BONE TYPE IS II, AND THEIR ORAL HYGIENE IS LISTED AS FAIR. THE PATIENT PRESENTED, ON (B)(6) 2022, FOR A PRIMARY PROCEDURE ON TOOTH #7. THE PATIENT RETURNED, ON (B)(6) 2023, BEFORE THE SECOND STAGE OF SURGERY WITH COMPLAINTS OF PAIN. THE PATIENT LATER RETURNED ON (B)(6) 2023, AND IT WAS DETERMINED THAT THE IMPLANT FAILED TO INTEGRATE, AND THE DEVICE WAS REMOVED AND NOT REPLACED. THE SYMPTOMS RESOLVED AFTER DEVICE REMOVAL AND THERE WAS NO PERMANENT PATIENT INJURY. NO ADDITIONAL MEDICAL/SURGICAL PROCEDURE WAS REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 72749 | HAHN TAPERED IMPLANT Ø3.5 X 11.5 MM | HAHN TAPERED IMPLANT | DZE | PRISMATIK DENTALCRAFT, INC. | 70-1154-IMP0006 | 6113527 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Female | Required Intervention |