CS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, JAPANESE,NON-UTS,DOM
Report
- Report Number
- 2249723-2023-02874
- Event Type
- Malfunction
- Date Received
- June 21, 2023
- Date of Event
- June 12, 2023
- Report Date
- April 3, 2024
- Manufacturer
- DATASCOPE CORP. - MAHWAH
- Product Code
- DSP
- UDI-DI
- 10607567108674
- PMA / PMN Number
- K031636
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION.
EVENT SITE POSTAL CODE AND STATE : (B)(6). IT WAS REPORTED THAT DURING THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) DEFECTIVE SCREEN DISPLAY. A GETINGE FIELD SERVICE ENGINEER (FSE) REPORT OF SCREEN DISPLAY FAILURE OCCURRENCE DATE UNKNOWN. THE SCREEN IS DISTORTED WHEN THE POWER IS TURNED ON. # 6/9 WHEN THE ENGINEER CHECKED, THERE WAS A SIMILAR SYMPTOM (THE SCREEN TURNED WHITE). TAP THE BACK OF THE MONITOR TO IMPROVE THE DISPLAY. THE VIDEO RECEIVER BOARD (D670-00-0736) AND FLAT CABLE WERE REPLACED (D012-00-1429), AND IT WAS RETURNED AFTER CONFIRMING THAT IT STARTED NORMALLY. THE DEVICE HAS BEEN DELIVERED TO THE CUSTOMER AND IS READY FOR CLINICAL USE. THERE WAS NO PATIENT INVOLVEMENT. THE DEFECTIVE COMPONENTS WERE RECEIVED FOR FURTHER INVESTIGATION. PLEASE REFER TO THE ROOT CAUSE EVALUATION FIELD FOR DETAILS. THE FAT PERFORMED A VISUAL INSPECTION AND FOUND THE PARTS TO BE IN GOOD CONDITION. THE FAT INSTALLED PN 0670-00-0736 INTO CS300 TEST FIXTURE SERIAL NUMBER (B)(6) AND TESTED THE PART TO FACTORY SPECIFICATIONS PER PROCEDURE NUMBER (B)(6) REVISION E AND THE CS300 SERVICE MANUAL PART NUMBER 0070-00-0689 REV W. NO ISSUE FOUND. INSTALLED PN 0012-00-1429 INTO CS300 TEST FIXTURE SERIAL NUMBER (B)(6) AND TESTED THE PART TO FACTORY SPECIFICATIONS PER PROCEDURE NUMBER (B)(4) REVISION E AND THE CS300 SERVICE MANUAL PART NUMBER 0070-00-0689 REV W. NO ISSUE FOUND. FAT COULD NOT VERIFY THE REPORTED FAILURE FOR ANY OF THESE PARTS. RETAINING THE PARTS IN THE FAILURE ANALYSIS AND TESTING DEPARTMENT PER PROCEDURE NUMBER (B)(4) REV. AP. THE NON-CONFORMANCES WITH THE RETURNED COMPONENTS WERE NOT CONFIRMED. HOWEVER, THE ROOT CAUSE OR THE MOST PROBABLE ROOT CAUSE IS NOT CONFIRMED.
IT WAS REPORTED THAT DURING START UP, THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) UNITS SCREEN S BLACK. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 207317 | CS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, JAPANESE,NON-UTS,DOM | SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL | DSP | DATASCOPE CORP. - MAHWAH | 0998-UC-3013-65 | 10607567108674 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |