I-STAT TROPONIN (CTNI) CARTRIDGE
Report
- Report Number
- 2245578-2023-00074
- Event Type
- Malfunction
- Date Received
- June 20, 2023
- Date of Event
- June 14, 2023
- Report Date
- July 7, 2023
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- MMI
- UDI-DI
- 10054749000187
- PMA / PMN Number
- K031739
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NZ
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT: # (B)(4).
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 28-JUN-2023. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AK (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). NO DEFICIENCY HAS BEEN IDENTIFIED FOR CTNI CARTRIDGE LOT B23044.
APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
ON 14-JUN-2023, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED A FALSE POSITIVE DISCREPANT RESULT OF 0.71 UG/L ON A PATIENT WITH CHEST PAIN. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD: DATE: COLLECTED: TESTED: RESULTS: I-STAT, (B)(6) 2023, 18:00, 18:06, 0.71. I-STAT, (B)(6) 2023, 18:00, 18:19, 0.02. I-STAT, (B)(6) 2023, 18:00, 18:32, 0.01. I-STAT, (B)(6) 2023, 18:00, 18:32, 0.01. I-STAT, (B)(6) 2023, 18:00, 18:34, 0.00. ALL I-STAT SAMPLES: 2ML LI HEP TUBE. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THE INVESTIGATION IS UNDERWAY. PER I-STAT SYSTEM MANUAL: ART: 715595-00V. REPORTABLE RANGE: 0.00 - 50.00 REFERENCE RANGE: 0.00 - 0.03 (REPRESENTS THE 0 TO 97.5% RANGE OF RESULTS). REFERENCE RANGE: 0.00 - 0.08 (REPRESENTS THE 0 TO 99% RANGE OF RESULTS).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1014213 | I-STAT TROPONIN (CTNI) CARTRIDGE | CTNI CARTRIDGE | MMI | ABBOTT POINT OF CARE | NA | B23044 | 10054749000187 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Female |