FDA Adverse Event Malfunction Summary report: N

HAMILTON MEDICAL AG

MDR report key: 17152177 · Received June 19, 2023

Report

Report Number
3001421318-2023-02412
Event Type
Malfunction
Date Received
June 19, 2023
Date of Event
October 18, 2023
Report Date
April 9, 2025
Manufacturer
HAMILTON MEDICAL AG
Product Code
CBK
PMA / PMN Number
K181216
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION ADDED IN FOLLOW-UP #1 (B)(4). THE ISSUE WAS DISCOVERED DURING START UP WITH NO PATIENT INVOLVED. CONSEQUENTLY, THE EVENT DID NOT CAUSE OR CONTRIBUTED TO A DEATH OR SERIOUS INJURY. THE ROOT CAUSE OF THE EVENT WAS DETERMINED TO BE A DEFECTIVE ESM BOARD. AFTER REPLACING THE ESM BOARD AND CALIBRATION OF THE DEVICE IT WAS FOUND TO BE FUNCTIONAL AND WAS RELEASED FOR USE. THE ISSUE IS CONSIDERED A REPORTABLE EVENT BECAUSE IT COULD POTENTIALLY IMPACT PATIENT SAFETY DURING VENTILATION SINCE THE DEFECT ON THE ESM BOARD COULD RESULT IN INADEQUATE VENTILATION SUPPORT. IN AGREEMENT WITH THE FDA, UDI NUMBERS (FIELD D4 IN THIS FORM) ARE ONLY PROVIDED FOR INCIDENTS THAT HAVE BEEN INITIALLY REPORTED BY HAMILTON MEDICAL AG SINCE OCTOBER 1ST, 2023.

Description of Event or Problem · 0

ACCORDING TO THE REPORT, THIS IS A CASE REPORT RECEIVED ON JUNE 10TH FROM HAMILTON MEDICAL INC. RELATED TO TECHNICAL FAULT OBSERVED ON A (B)(6) 2022 ON A HAMILTON C1 VENTILATOR (SN# (B)(6), SOFTWARE VERSION 2.2.10). ACCORDING TO THE REPORTER, THERE IS NO PATIENT INVOLVED. THE ISSUE DID NOT OCCURRED DURIGN VENTILATION. PRELIMINARY ANALYSIS REVEALED THAT: IN THE CONFIGURATION MENU THE OPTIONS DISAPPEARED. OPTIONS ARE NO LONGER AVAILABLE. IN THE CONFIGURATION MENU THERE ARE NO OPTIONS DISPLAYED IN "SW OPTIONS". 'REALTIME CLOCK FAILURE' MIGHT BE DISPLAYED AS WELL. OR OPTION KEYS ARE NOT VISIBLE / AVAILABLE ANYMORE. VENTILATION CANNOT START. POTENTIAL ROOT CAUSE IS RELATED TO THE RESET OF THE REALTIME CLOCK (RTC) DUE TO LOSS OF SUPPLY VOLTAGE FOR LONGER THAN APPROX. ONE WEEK ESM BOARD NOT PROPERLY CONNECTED. AS IMMEDIATE CORRECTION, THE FOLLOWING ACTION HAVE BEEN PERFORMED: SET CORRECT REAL TIME CLOCK (RTC) IN SERVICE SOFTWARE PAGE 1401 AND RESTART THE UNIT. INSTALL THE LATEST SOFTWARE VERSION. ENTER THE ADULT ACTIVATION CODE AND RESTART THE UNIT. IF FAILURE PERSISTS ALSO CHECK THE ESM BOARD FOR PROPER CONNECTION TO THE MAINBOARD.

Description of Event or Problem · 0

ACCORDING TO THE REPORT, THIS IS A CASE REPORT RECEIVED ON JUNE 10TH FROM HAMILTON MEDICAL INC. RELATED TO TECHNICAL FAULT OBSERVED ON OCTOBER 18TH 2022 ON A HAMILTON C1 VENTILATOR (SN# (B)(6), SOFTWARE VERSION 2.2.10) ACCORDING TO THE REPORTER, THERE IS NO PATIENT INVOLVED. THE ISSUE DID NOT OCCURRED DURIGN VENTILATION. PRELIMINARY ANALYSIS REVEALED THAT: IN THE CONFIGURATION MENU THE OPTIONS DISAPPEARED. - OPTIONS ARE NO LONGER AVAILABLE. IN THE CONFIGURATION MENU THERE ARE NO OPTIONS DISPLAYED IN "SW OPTIONS". 'REALTIME CLOCK FAILURE' MIGHT BE DISPLAYED AS WELL. OR OPTION KEYS ARE NOT VISIBLE / AVAILABLE ANYMORE. VENTILATION CANNOT START. POTENTIAL ROOT CAUSE IS RELATED TO THE RESET OF THE REALTIME CLOCK (RTC) DUE TO LOSS OF SUPPLY VOLTAGE FOR LONGER THAN APPROX. ONE WEEK ESM BOARD NOT PROPERLY CONNECTED. AS IMMEDIATE CORRECTION, THE FOLLOWING ACTION HAVE BEEN PERFORMED: SET CORRECT REAL TIME CLOCK (RTC) IN SERVICE SOFTWARE PAGE 1401 AND RESTART THE UNIT INSTALL THE LATEST SOFTWARE VERSION. ENTER THE ADULT ACTIVATION CODE AND RESTART THE UNIT IF FAILURE PERSISTS ALSO CHECK THE ESM BOARD FOR PROPER CONNECTION TO THE MAINBOARD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
778127 HAMILTON MEDICAL AG HAMILTON-C1 CBK HAMILTON MEDICAL AG HAMILTON-C1

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown