FDA Adverse Event Malfunction Summary report: N

TELESCOPE "TRUEVIEW II", 4 MM, 30°, AUTOCLAVABLE

MDR report key: 17148677 · Received June 16, 2023

Report

Report Number
9610773-2023-01673
Event Type
Malfunction
Date Received
June 16, 2023
Report Date
August 3, 2023
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HRX
UDI-DI
04042761010542
PMA / PMN Number
K951354
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION AND CORRECTION TO G2. G2 - CHECKED "OTHER" AND ADDED THE COUNTRY FRANCE. BASED ON THE RESULTS OF THE INVESTIGATION, SINCE NO DEVICE MALFUNCTION WAS CONFIRMED DURING EVALUATION, THE DEFINITIVE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED.

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION AND REPAIR. DURING THE EVALUATION, THE MALFUNCTION INITIALLY REPORTED COULD NOT BE CONFIRMED. DISTORTED IMAGE IDENTIFIED CAUSED BY DAMAGE LENS WITHOUT COMPONENT BREAKAGE. THE INVESTIGATION IS STILL IN PROGRESS; THEREFORE, THE ROOT CAUSE OF THE REPORTED DEFECT CANNOT BE DETERMINED AT THIS TIME. HOWEVER, IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED ACCORDINGLY. IN GENERAL, THE CUSTOMER IS REQUIRED TO INSPECT THE DEVICE FOR DEFECTS, CHECK THE FUNCTION OF ALL DEVICES AND HAVE ALTERNATE EQUIPMENT PRIOR TO USE. A MANUFACTURING AND QUALITY CONTROL REVIEW WAS PERFORMED FOR THE AFFECTED SERIAL NUMBER WITHOUT SHOWING ANY NON-CONFORMITIES OR DEVIATIONS REGARDING THE DESCRIBED ISSUES. OLYMPUS WILL CONTINUE TO MONITOR THE FIELD PERFORMANCE OF THIS DEVICE. ADDITIONAL PMA/510(K): K944072 / K950076.

Description of Event or Problem · 0

IT WAS REPORTED TO OLYMPUS THAT A TELESCOPE HAD THE DISTAL LENS BROKEN. IT IS UNKNOWN WHEN THE REPORTED ISSUE WAS FOUND. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT REPORTED TO OLYMPUS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
545424 TELESCOPE "TRUEVIEW II", 4 MM, 30°, AUTOCLAVABLE RIGID SCOPE HRX OLYMPUS WINTER & IBE GMBH A70941A 619280 04042761010542

Patients

Seq Age Sex Outcome Treatment
1 Unknown