FDA Adverse Event Malfunction Summary report: N

MENTOR SMOOTH ROUND HIGH PROFILE

MDR report key: 17146874 · Received June 16, 2023

Report

Report Number
1645337-2023-07209
Event Type
Malfunction
Date Received
June 16, 2023
Date of Event
May 23, 2023
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001645
PMA / PMN Number
P990075
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ON JUN 22, 2023, THE MENTOR FAILURE ANALYSIS LAB RECEIVED A DEVICE (LOT NUMBER 9830880) FOR EVALUATION. ON JUL 3, 2023, IT WAS CONFIRMED THIS WAS THE IMPACTED PRODUCT. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. MANUFACTURER¿S REFERENCE NUMBER: (B)(4)

Additional Manufacturer Narrative · 0

DEVICE EVALUATION SUMMARY: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR THEN CONDUCTED A VISUAL INSPECTION, LEAK TESTING, AND MICROSCOPIC EXAMINATION OF THE RETURNED DEVICE. DURING A VISUAL EVALUATION, NO APPARENT DAMAGE OR VISUAL ANOMALIES WERE OBSERVED ON THE SM HPS DV 330CC RETURNED DEVICE. LEAK TESTING WAS PERFORMED, IN ACCORDANCE WITH MENTOR PROCEDURES, AND IT IDENTIFIED A TEAR ON THE ANTERIOR VIEW, MEASURING APPROXIMATELY 0.5 CM. MICROSCOPIC EXAMINATION WAS PERFORMED AND THE CAUSE OF THE DEFLATION COULD NOT BE IDENTIFIED. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. BASED ON THE INFORMATION CURRENTLY AVAILABLE, A PRODUCT ISSUE WAS IDENTIFIED DURING THE INVESTIGATION OF THE SAMPLE RECEIVED. THIS PRODUCT ISSUE WILL BE ADDRESSED THROUGH MENTOR¿S QUALITY SYSTEM. ACCORDING TO THE MANUFACTURE'S INVESTIGATION, THE PROCESS CONTROLS AND DATA RECORDS COLLECTED FOR THE DEVICE ARE WITHIN SPECIFICATION. TEAR REPORTED ON PRODUCT COMPLAINT IS NOT ABLE TO BE CONFIRMED AS MANUFACTURING DEFECT. AS PART OF MENTOR¿S QUALITY PROCESS ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A SALINE MENTOR SMOOTH ROUND HIGH PROFILE 330CC BREAST IMPLANT BEING USED IN A BREAST IMPLANTATION PROCEDURE WAS FOUND TO BE DEFLATED DURING THE OPERATION. NO ADVERSE EVENT WAS EXPERIENCED BY THE PATIENT. THE SURGEON USED BACKUP DEVICES, AND THERE WERE NO REPORTED PATIENT CONSEQUENCES OR PROCEDURAL DELAYS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
462328 MENTOR SMOOTH ROUND HIGH PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 3503330 9830880 00081317001645

Patients

Seq Age Sex Outcome Treatment
1 Unknown