FDA Adverse Event Other Summary report: N

HUDSON CONCHATHERM NEPTUNE

MDR report key: 1714583 · Received June 9, 2010

Report

Report Number
3003898360-2010-00236
Event Type
Other
Date Received
June 9, 2010
Date of Event
May 11, 2010
Report Date
May 12, 2010
Manufacturer
TELEFLEX MEDICAL
Product Code
BTT
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TYPE OF REPORTABLE EVENT - OTHER - THE WATER TUBING WAS CLAMPED OFF AND PATIENT DID NOT RECEIVE HUMIDIFICATION FOR 14 HOURS. THE PATIENT (B)(6) REQUIRED EXTENSIVE SUCTIONING AND LAVAGE. EVALUATION, METHOD: OTHER - HISTORICAL INFORMATION REVIEW. RESULTS: OTHER - HISTORICAL INFORMATION REVIEW FROM (B)(6) 2009 TO (B)(6) 2010 SHOWED NO SIMILAR COMPLAINTS FOR THIS MATERIAL NUMBER. NO DEVICE SAMPLE AVAILABLE FOR EVALUATION. NO LOT NUMBER PROVIDED. ROOT CAUSE - CUSTOMER MISUSE.

Description of Event or Problem · 1

THE EVENT IS REPORTED AS: WHILE INSERTING A NEW WATER BOTTLE, THE WATER TUBING WAS CLAMPED OFF AND THIS TUBING WAS NOT UN-CLAMPED. THE PATIENT (B)(6) WENT WITHOUT HUMIDITY FOR 14 HOURS. THE PATIENT DECOMPENSATED GOING FROM THE BABYLOG VENTILATOR TO OSCILLATOR SUPPORT AND INCREASED FI02. THIS PATIENT ALSO REQUIRED EXTENSIVE SUCTIONING AND LAVAGE. THE CUSTOMER REPORTS THAT THE THERAPIST DID NOT NOTICE THE CLAMPED TUBING UNTIL A LATER TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HUDSON CONCHATHERM NEPTUNE HEATED HUMIFICATION DEVICE BTT TELEFLEX MEDICAL NA UNK

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention