FDA Adverse Event Injury Summary report: N

BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE

MDR report key: 17145083 · Received June 16, 2023

Report

Report Number
1024879-2023-00385
Event Type
Injury
Date Received
June 16, 2023
Date of Event
May 23, 2023
Report Date
August 2, 2023
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
FMI
UDI-DI
50382903686075
PMA / PMN Number
K982541
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 20 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING, EACH HAVING THEIR ECLIPSE SHIELDS ACTIVATED, AND NO ISSUES WERE OBSERVED RELATING TO DEFECTIVE LOCKING MECHANISM AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE DEFECTIVE LOCKING MECHANISM. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.

Additional Manufacturer Narrative · 0

B.3. DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE THE CAP CONSTANTLY FALLS OFF. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "CUSTOMER PROBLEM: CUSTOMER REPORTS AN ISSUE WITH LOT#: 2090159. CUSTOMER REPORTS THE SHIELDS FROM ECLIPSE NEEDLES CONSTANTLY FALL OFF."

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE THE CAP CONSTANTLY FALLS OFF. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "CUSTOMER PROBLEM: CUSTOMER REPORTS AN ISSUE WITH LOT#: 2090159. CUSTOMER REPORTS THE SHIELDS FROM ECLIPSE NEEDLES CONSTANTLY FALL OFF."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
692090 BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON, DICKINSON & CO., (BD) 368607 2090159 50382903686075

Patients

Seq Age Sex Outcome Treatment
1 Unknown