FDA Adverse Event Malfunction Summary report: N

VISERA CYSTO-NEPHRO VIDEOSCOPE

MDR report key: 17144814 · Received June 16, 2023

Report

Report Number
3002808148-2023-06079
Event Type
Malfunction
Date Received
June 16, 2023
Date of Event
May 19, 2023
Report Date
July 20, 2023
Manufacturer
SHIRAKAWA OLYMPUS CO., LTD.
Product Code
FAJ
UDI-DI
04953170411199
PMA / PMN Number
K221683
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. THE G2 FIELD WAS CORRECTED BASED ON INFORMATION AVAILABLE AT THE INITIAL REPORT SUBMISSION. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. IT HAS BEEN OVER 7 YEARS SINCE THE SUBJECT DEVICE WAS MANUFACTURED. BASED ON THE RESULTS OF THE INVESTIGATION, A DEFINITIVE ROOT CAUSE COULD NOT BE ESTABLISHED. THE EVENT CAN BE PREVENTED BY FOLLOWING THE INSTRUCTIONS FOR USE WHICH STATES: "WARNING - DO NOT STRIKE, BEND, HIT, PULL, TWIST, OR DROP THE ENDOSCOPE¿S DISTAL END, INSERTION TUBE, BENDING SECTION, AND CONTROL SECTION WITH EXCESSIVE FORCE. THE ENDOSCOPE MAY BE DAMAGED AND COULD CAUSE PATIENT INJURY, SUCH AS BURNS, BLEEDING, AND/OR PERFORATION. IT COULD ALSO CAUSE PARTS OF THE ENDOSCOPE TO FALL OFF INSIDE THE PATIENT." OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION AND THE CUSTOMER'S ALLEGATION WAS ABLE TO BE CONFIRMED. DURING THE DEVICE EVALUATION IT WAS OBSERVED THAT THE BLACK COVERING ON THE TIP WAS BLASTED OFF AND THE BENDING SECTION COVER WAS TORN/ CUT. DUE TO THE TORN BENDING SECTION COVER, THE SCOPE DID NOT PASS THE LEAK TEST. ADDITIONALLY, THE PLASTIC DISTAL END COVER INSULATION TEST WAS UNABLE TO BE PERFORMED DUE TO THE TORN BENDING SECTION COVER. LASTLY, IT WAS OBSERVED THAT THERE WAS CORROSION AROUND THE OBJECTIVE LENS. THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 0

THE CUSTOMER REPORTED TO OLYMPUS THE VISERA CYSTO-NEPHRO VIDEOSCOPE BENDING SECTION COVER FELL OFF AND THE BLACK COVERING AT THE TIP WAS BLASTED OFF THE DISTAL END. THE REPORTED ISSUE OCCURRED AT REPROCESSING DURING STERILE PROCESSING. THERE WAS NO PATIENT/USER HARM OR INJURY REPORTED DUE TO THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
661807 VISERA CYSTO-NEPHRO VIDEOSCOPE CYSTO-NEPHRO VIDEOSCOPE FAJ SHIRAKAWA OLYMPUS CO., LTD. CYF-V2 04953170411199

Patients

Seq Age Sex Outcome Treatment
1 Unknown