DUAL 4 EXTENSION
Report
- Report Number
- 1627487-2010-00059
- Event Type
- Malfunction
- Date Received
- February 26, 2010
- Date of Event
- January 27, 2010
- Report Date
- January 28, 2010
- Manufacturer
- ADVANCED NEUROMODULATION SYSTEMS, INC.
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
EVAL METHOD: THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: THE DEVICE HISTORY AND STERILIZATION RECORDS REVIEWED WERE FOUND TO MEET SPECS AND NO ANOMALIES WERE FOUND. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. ANS HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. ANS DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 3 OF 4. (REFER TO MFR'S REPORT NUMBERS, 1627487-2010-0053 FOR DEVICE 1, 1627487-2010-0058 FOR DEVICE 2 AND 1627487-2010-00107 FOR DEVICE 4). IT WAS REPORTED, THAT DURING A LEAD REVISION SURGERY (DUE TO LEAD MIGRATION) THE LEADS CAME OUT OF THE IPG HEADER. THE DOCTOR TIGHTENED THE SCREW WHILE USING A WRENCH, BUT THE LEADS CONTINUED TO SLIP OUT OF THE DEVICE. THE LEADS FUNCTIONED PROPERLY BUT WERE DISCARDED AFTER THE FAILED ATTEMPT TO TIGHTEN THEM INTO THE IPG.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DUAL 4 EXTENSION | SPINAL CORD STIMULATION LEAD | LGW | ADVANCED NEUROMODULATION SYSTEMS, INC. | 3341 | 2753958 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR | Other |