FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET

MDR report key: 17139787 · Received June 15, 2023

Report

Report Number
1024879-2023-00380
Event Type
Malfunction
Date Received
June 15, 2023
Date of Event
May 4, 2023
Report Date
June 26, 2023
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
JKA
UDI-DI
50382903673440
PMA / PMN Number
K030573
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: MATERIAL #: 367344 LOT/BATCH #: 3061934 BD HAD NOT RECEIVED SAMPLES, BUT 1 PHOTO WAS PROVIDED FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND IT SHOWS THE PRODUCT PACKAGING. ADDITIONALLY, 30 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL DRAW TESTING AND RETRACTION TESTING AND NO ISSUES WERE OBSERVED RELATING TO LEAKAGE / SPLATTER AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE LEAKAGE. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET REPORTED THAT GREEN & BLUE BUTTERFLY PUSH BUTTON NEEDLES SPRAY TINY BLOOD DROPLETS WHEN THE PUSH BUTTON SAFETY IS ACTIVATED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "IT WAS REPORTED BY THE CUSTOMER THAT 367344_3061934 LEAKING."

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET REPORTED THAT GREEN & BLUE BUTTERFLY PUSH BUTTON NEEDLES SPRAY TINY BLOOD DROPLETS WHEN THE PUSH BUTTON SAFETY IS ACTIVATED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "IT WAS REPORTED BY THE CUSTOMER THAT (B)(6)LEAKING."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2103042 BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET JKA BECTON, DICKINSON & CO., (BD) 367344 3061934 50382903673440

Patients

Seq Age Sex Outcome Treatment
1 Unknown