FDA Adverse Event Malfunction Summary report: N

ADULT DUAL HEATED BREATHING CIRCUIT

MDR report key: 1713881 · Received May 27, 2010

Report

Report Number
9611451-2010-00294
Event Type
Malfunction
Date Received
May 27, 2010
Date of Event
April 20, 2010
Report Date
April 23, 2010
Manufacturer
FISHER & PAYKEL HEALTHCARE, LTD.
Product Code
BTT
PMA / PMN Number
K983112
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE RT200 IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. METHOD: THE RETURNED BREATHING CIRCUIT WAS TESTED FOR BENT PINS. RESULTS: THE INSPIRATORY LIMB HEATER WIRE PIN ON THE BREATHING CIRCUIT WAS BENT, PREVENTING FULL INSERTION OF A HEATER WIRE ADAPTOR. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 091209. CONCLUSION: ALL BREATHING CIRCUITS ARE TESTED FOR CONNECTIVITY AND ELECTRICAL CONTINUITY DURING PRODUCTION AND THOSE THAT FAILED ARE REJECTED. IT IS POSSIBLE FOR THE USER TO BEND THE HEATER WIRE PINS DURING USE IF THE HEATER WIRE ADAPTOR IS INSERTED INTO THE HEATER WIRE PLUG ON AN ANGLE. FOR BENT PINS REPORTED TO US BY HEALTH CARE FACILITIES, IT IS IMPOSSIBLE FOR US TO DETERMINE WHETHER THE PINS WERE BENT DURING PRODUCTION OR BY THE END USER. DAMAGED PINS DO NOT PRECLUDE VENTILATION OF THE PATIENT, BUT PREVENT THE HEATING OF THE GAS DELIVERED. (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT AN RT200 ADULT DUAL HEATED BREATHING CIRCUIT HAD BENT PINS. THE HOSPITAL NOTICED THAT THE HEATER WIRE PINS WERE BENT PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADULT DUAL HEATED BREATHING CIRCUIT BTT FISHER & PAYKEL HEALTHCARE, LTD. RT200 091209

Patients

Seq Age Sex Outcome Treatment
1