FDA Adverse Event Malfunction Summary report: N

PUMP 1886 780G OUS BLE PUMP MG/DL

MDR report key: 17132422 · Received June 15, 2023

Report

Report Number
2032227-2023-217889
Event Type
Malfunction
Date Received
June 15, 2023
Date of Event
May 22, 2023
Report Date
August 21, 2023
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

PUMP WAS RECEIVED WITH A CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. UNABLE TO PERFORM THE DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, SLEEP CURRENT MEASUREMENT TEST, ACTIVE CURRENT MEASUREMENT TEST AND SELF TEST DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND CORRODED PCBA 1, PCBA 2 AND FORCE SENSOR. MOISTURE DAMAGE ON THE MOTOR. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUMP. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM WAS TRIGGERED BY A PUMP ERROR 35 FATAL ALARM CONFIRMED IN THE HISTORY FILE ON 05/22/2023 20:34:03.000 AND ON 05/22/2023 20:44:00.000 DUE TO CORRODED FORCE SENSOR. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: MINOR SCRATCHED DISPLAY WIDOW, MISSING DISPLAY WINDOW COVER, SCRATCHED CASE, CRACKED BATTERY TUBE THREADS, CRACKED CASE BATTERY TUBE SIDE, CRACKED CASE AT THE CORNER OF THE BELT CLIP RAIL, STAINED SERIAL NUMBER LABEL, FADED END CAP ADDRESS LABEL, PILLOWING KEYPAD OVERLAY, AND CRACKED KEYPAD OVERLAY AT THE SELECT BUTTON. THE PUMP DID NOT HAVE A BATTERY INSTALLED WHEN RECEIVED. NO DAMAGE NOTED ON THE ORIGINAL BATTERY CAP. PLEASE SEE BELOW FOR THE PUMP ERRORS/ALARMS NOTED 2 DAYS PRIOR TO THE EVENT DATE 22-MAY-2023 IN THE PUMP HISTORY FILE. 05/20/2023 00:39:00.000 ALARMALERTNOTIFICATION (40), FAULTNUMBER: LOSTSENSOR1ALERT (780) 05/22/2023 20:34:03.000 ALARMALERTNOTIFICATION (40), FAULTNUMBER: BROKENFORCESENSORERROR (35) 05/22/2023 20:44:00.000 ALARMALERTNOTIFICATION (40), FAULTNUMBER: BROKENFORCESENSORERROR (35) 05/22/2023 20:56:11.000 BATTERYREMOVED (55) 05/22/2023 20:56:11.000 ALARMALERTNOTIFICATION (40), FAULTNUMBER: BATTERYREMOVED (84). UNABLE TO TEST FOR LOST SENSOR ALARM DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. PUMP ERROR 35 WAS CONFIRMED IN THE PUMP HISTORY FILE. LOSTSENSOR1ALERT (780) UNKNOWN. PUMP ERROR 35 - BROKENFORCESENSORERROR (35) CONFIRMED. UNABLE TO PERFORM THE DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, SLEEP CURRENT MEASUREMENT TEST, ACTIVE CURRENT MEASUREMENT TEST AND SELF TEST DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM CONFIRMED DUE TO PUMP ERROR 35 ALARM. PUMP ERROR 35 ALARM CONFIRMED DUE TO CORRODED FORCE SENSOR. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE NGP 780G INSULIN PUMP WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE NGP INSULIN PUMP, WHICH IS MARKETED IN THE UNITED STATES MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER RECEIVED A CRITICAL PUMP ERROR ON THE INSULIN PUMP SCREEN. TROUBLESHOOTING WAS PERFORMED AND FOUND THAT THE CUSTOMER RECEIVED AN OPEN BOOK IMAGE ALARM. IT WAS REPORTED THAT THE INSULIN PUMP PERFORMED SAFETY CHECKS AND THE ERROR WAS FOUND AND THE ISSUE WAS NOT RESOLVED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE THE USE OF THE INSULIN PUMP. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1408774 PUMP 1886 780G OUS BLE PUMP MG/DL AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1886 HG526V9ZZ

Patients

Seq Age Sex Outcome Treatment
1 Unknown