FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES

MDR report key: 17130951 · Received June 14, 2023

Report

Report Number
9617032-2023-00713
Event Type
Malfunction
Date Received
June 14, 2023
Date of Event
May 22, 2023
Report Date
May 30, 2023
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
PMA / PMN Number
BK050036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY: MATERIAL #: 365301 LOT/BATCH #: 2192190 BD HAD NOT RECEIVED SAMPLES, BUT ONE PHOTO WAS PROVIDED FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND THE INDICATED FAILURE MODE FOR INCORRECT CAP COLOR WAS OBSERVED. ADDITIONALLY, ONE HUNDRED (100) RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND THE ISSUE OF INCORRECT CAP COLOR WAS NOT OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING THE BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES THAT THE STOPPER WAS A DIFFERENT COLOR. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ONE TUBE OUT OF A WHOLE TRAY CAME WITH YELLOW CAP INSTEAD OF CLEAR CAP - ELIEZER CLINIC THE TUBE HAD A YELLOW CAP , THE CLINIC DIDN'T SAVE THE TUBE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1875336 BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 2192190

Patients

Seq Age Sex Outcome Treatment
1 Unknown