ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO
Report
- Report Number
- 2953200-2010-01076
- Event Type
- Death
- Date Received
- June 4, 2010
- Date of Event
- May 19, 2010
- Report Date
- May 19, 2010
- Manufacturer
- MEDTRONIC CARDIOVASCULAR GALWAY
- Product Code
- MIH
- PMA / PMN Number
- P990020
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B) (4). EVALUATION, RESULTS: DEATH, RUPTURE. LARGE ANEURYSM AND ANGULATED AORTIC NECK. PT WAS LOST TO FOLLOW-UP. EVALUATION, CONCLUSIONS: LARGE ANEURYSM AND ANGULATED AORTIC NECK. PT WAS LOST TO FOLLOW-UP.
AN ANEURX STENT GRAFT WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 10.9 CM ABDOMINAL AORTIC ANEURYSM APPROXIMATELY 4 YEARS AGO. THE AORTIC NECK WAS 27 MM IN DIAMETER WITH SOME ANGULATION. THE PT WAS LOST TO FOLLOW-UP BUT RETURNED TO THE HOSPITAL 6 MONTHS AGO. IT WAS NOTED THAT THE BIFURCATED STENT GRAFT WAS LOCATED SOME UNK DISTANCE BELOW THE RENAL ARTERIES AND IT MAY HAVE BEEN PLACED LOW AT THE TIME OF IMPLANT; HOWEVER, THERE WAS NO ENDOLEAK AND NO REPORTED MIGRATION. THE ANEURYSM WAS LARGE AT 10.9 CM. THE DECISION WAS MADE TO BUILD THE BIFURCATED STENT GRAFT UP TO THE LEVEL OF THE RENAL ARTERIES WITH 2 ANEURX 28 MM CUFFS AND A TALENT 34 MM CUFF (SEE MFR # 2953200-2010-01075). THE FIRST ANEURX CUFF WAS PLACED AT THE GATE OF THE BIFUR AND AT THE 6 MONTH CT FOLLOW-UP, THE ANEURYSM DIAMETER WAS 10.1 CM WITH NO ENDOLEAK VISIBLE. FIVE DAYS LATER, THE PT PRESENTED EMERGENTLY WITH FLANK PAIN; HOWEVER, THERE WAS NO ENDOLEAK OR RUPTURE ON CT. THE PT WAS THOUGHT TO BE STABLE AND WAS MOVED TO THE ICU, BUT EXPIRED A SHORT TIME LATER FROM A RUPTURED ANEURYSM. IT IS LIKELY THAT THERE WAS A SMALL HOLE (TYPE 3 ENDOLEAK) MAY BE PRESENT ABOVE THE GATE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO | MIH | MEDTRONIC CARDIOVASCULAR GALWAY | NA | V00363579 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Death |