FDA Adverse Event Death Summary report: N

ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO

MDR report key: 1712884 · Received June 4, 2010

Report

Report Number
2953200-2010-01076
Event Type
Death
Date Received
June 4, 2010
Date of Event
May 19, 2010
Report Date
May 19, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P990020
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B) (4). EVALUATION, RESULTS: DEATH, RUPTURE. LARGE ANEURYSM AND ANGULATED AORTIC NECK. PT WAS LOST TO FOLLOW-UP. EVALUATION, CONCLUSIONS: LARGE ANEURYSM AND ANGULATED AORTIC NECK. PT WAS LOST TO FOLLOW-UP.

Description of Event or Problem · 1

AN ANEURX STENT GRAFT WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 10.9 CM ABDOMINAL AORTIC ANEURYSM APPROXIMATELY 4 YEARS AGO. THE AORTIC NECK WAS 27 MM IN DIAMETER WITH SOME ANGULATION. THE PT WAS LOST TO FOLLOW-UP BUT RETURNED TO THE HOSPITAL 6 MONTHS AGO. IT WAS NOTED THAT THE BIFURCATED STENT GRAFT WAS LOCATED SOME UNK DISTANCE BELOW THE RENAL ARTERIES AND IT MAY HAVE BEEN PLACED LOW AT THE TIME OF IMPLANT; HOWEVER, THERE WAS NO ENDOLEAK AND NO REPORTED MIGRATION. THE ANEURYSM WAS LARGE AT 10.9 CM. THE DECISION WAS MADE TO BUILD THE BIFURCATED STENT GRAFT UP TO THE LEVEL OF THE RENAL ARTERIES WITH 2 ANEURX 28 MM CUFFS AND A TALENT 34 MM CUFF (SEE MFR # 2953200-2010-01075). THE FIRST ANEURX CUFF WAS PLACED AT THE GATE OF THE BIFUR AND AT THE 6 MONTH CT FOLLOW-UP, THE ANEURYSM DIAMETER WAS 10.1 CM WITH NO ENDOLEAK VISIBLE. FIVE DAYS LATER, THE PT PRESENTED EMERGENTLY WITH FLANK PAIN; HOWEVER, THERE WAS NO ENDOLEAK OR RUPTURE ON CT. THE PT WAS THOUGHT TO BE STABLE AND WAS MOVED TO THE ICU, BUT EXPIRED A SHORT TIME LATER FROM A RUPTURED ANEURYSM. IT IS LIKELY THAT THERE WAS A SMALL HOLE (TYPE 3 ENDOLEAK) MAY BE PRESENT ABOVE THE GATE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00363579

Patients

Seq Age Sex Outcome Treatment
1 73 YR Death