MPS DELIVERY SET
Report
- Report Number
- 1649914-2023-00018
- Event Type
- Malfunction
- Date Received
- June 14, 2023
- Date of Event
- May 15, 2023
- Report Date
- June 14, 2023
- Manufacturer
- QUEST MEDICAL, INC.
- Product Code
- DTR
- UDI-DI
- 00634624501126
- PMA / PMN Number
- K953838
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE DEVICE WAS RECEIVED BY QUEST MEDICAL, INC. ON 05/26/2023. VISUAL INSPECTION CONFIRMED THE PRESENCE OF BLOOD ON THE TOP FLANGE OF THE HEAT EXCHANGER. IT WAS ALSO NOTED THAT THE BELLOW WITHIN THE HEX HAD A DEEP INDENTATION IN ITS NECK WHERE THE BLOOD POOLED. THE ROOT CAUSE WAS INFERRED TO BE DUE TO THE DEEP NECK DURING MOLDING BY THE SUPPLIER, LEADING TO INADEQUATE GLUE APPLICATION TO PREVENT LEAKAGE. QUEST WILL CONTINUE TO MONITOR THIS COMPLAINT CONDITION. QUEST HAS CONCLUDED ITS INVESTIGATION.
THE SAMPLE HAS BEEN RECEIVED, BUT ANALYSIS OF THE SAMPLE HAS NOT YET BEEN COMPLETED TO DATE. QUEST WILL FILE A SUPPLEMENTAL REPORT AT THE CONCLUSION OF THE INVESTIGATION.
IT WAS REPORTED TO QUEST MEDICAL BY (B)(6) OF AN ALLEGED ISSUE WITH PRODUCT CODE 5001102. IN THE REPORT, IT WAS STATED THAT BLOOD WAS ALLEGEDLY OBSERVED IN THE HEAT EXCHANGER. IT WAS CONFIRMED THAT THERE WERE NO PATIENT COMPLICATIONS. LOT NUMBER OF THE DEVICE WAS NOT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1230422 | MPS DELIVERY SET | CPBP HEAT EXCHANGER | DTR | QUEST MEDICAL, INC. | 5001102 | 00634624501126 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |