FDA Adverse Event Malfunction Summary report: N

MPS DELIVERY SET

MDR report key: 17128549 · Received June 14, 2023

Report

Report Number
1649914-2023-00018
Event Type
Malfunction
Date Received
June 14, 2023
Date of Event
May 15, 2023
Report Date
June 14, 2023
Manufacturer
QUEST MEDICAL, INC.
Product Code
DTR
UDI-DI
00634624501126
PMA / PMN Number
K953838
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RECEIVED BY QUEST MEDICAL, INC. ON 05/26/2023. VISUAL INSPECTION CONFIRMED THE PRESENCE OF BLOOD ON THE TOP FLANGE OF THE HEAT EXCHANGER. IT WAS ALSO NOTED THAT THE BELLOW WITHIN THE HEX HAD A DEEP INDENTATION IN ITS NECK WHERE THE BLOOD POOLED. THE ROOT CAUSE WAS INFERRED TO BE DUE TO THE DEEP NECK DURING MOLDING BY THE SUPPLIER, LEADING TO INADEQUATE GLUE APPLICATION TO PREVENT LEAKAGE. QUEST WILL CONTINUE TO MONITOR THIS COMPLAINT CONDITION. QUEST HAS CONCLUDED ITS INVESTIGATION.

Additional Manufacturer Narrative · 0

THE SAMPLE HAS BEEN RECEIVED, BUT ANALYSIS OF THE SAMPLE HAS NOT YET BEEN COMPLETED TO DATE. QUEST WILL FILE A SUPPLEMENTAL REPORT AT THE CONCLUSION OF THE INVESTIGATION.

Description of Event or Problem · 0

IT WAS REPORTED TO QUEST MEDICAL BY (B)(6) OF AN ALLEGED ISSUE WITH PRODUCT CODE 5001102. IN THE REPORT, IT WAS STATED THAT BLOOD WAS ALLEGEDLY OBSERVED IN THE HEAT EXCHANGER. IT WAS CONFIRMED THAT THERE WERE NO PATIENT COMPLICATIONS. LOT NUMBER OF THE DEVICE WAS NOT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1230422 MPS DELIVERY SET CPBP HEAT EXCHANGER DTR QUEST MEDICAL, INC. 5001102 00634624501126

Patients

Seq Age Sex Outcome Treatment
1 Unknown