FDA Adverse Event
Other
Summary report: N
HALF POWER SUPERSTAND STANDUP WHEELCHAIR
MDR report key: 1712829
·
Received June 5, 2010
Report
- Report Number
- 3007494904-2008-00001
- Event Type
- Other
- Date Received
- June 5, 2010
- Date of Event
- June 4, 2008
- Report Date
- June 2, 2010
- Manufacturer
- THE STANDING COMPANY
- Product Code
- IPL
- PMA / PMN Number
- K032101
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE WAS RETURNED TO OUR FACILITY ON 06/12/2008. DEVICE WAS INSPECTED. RECREATION OF EVENT ON (B) (6) 2008 AT DEVICE MFG FACILITY. UPON INVESTIGATION, IT WAS NOTED THAT THE SET SCREW ON THE ARM SWIVEL HOUSING, WHICH IS LOCTITED, THE LOCTITE DOES NOT ADHERE PROPERLY ONCE ARM ASSEMBLY IS INSTALLED WITH GREASE. DEVICE WAS REPLACED WITH NEW ARM ASSEMBLY. CORRECTIVE ACTION WAS GENERATED FROM INVESTIGATION. DEVICE WAS RETURNED TO PT (B) (6) IN (B) (6) 2008, WHERE HE WAS ABLE TO CONTINUE USE OF THE CHAIR. ISSUE WAS RESOLVED.
Description of Event or Problem · 1
PT'S MOM CALLED IN AND SAID THAT HER SON, PT: (B) (6), FELL OUT OF HIS STANDING WHEELCHAIR (MODEL: HPS-2, (B) (4)), AND HIT HIS HEAD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HALF POWER SUPERSTAND STANDUP WHEELCHAIR | STANDUP WHEELCHAIR | IPL | THE STANDING COMPANY | HPS-2 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR | Other | HPS-2, 2 YRS OF USE| HALF POWER SUPERSTAND STANDUP WHEELCHAIR |