FDA Adverse Event Other Summary report: N

HALF POWER SUPERSTAND STANDUP WHEELCHAIR

MDR report key: 1712829 · Received June 5, 2010

Report

Report Number
3007494904-2008-00001
Event Type
Other
Date Received
June 5, 2010
Date of Event
June 4, 2008
Report Date
June 2, 2010
Manufacturer
THE STANDING COMPANY
Product Code
IPL
PMA / PMN Number
K032101
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS RETURNED TO OUR FACILITY ON 06/12/2008. DEVICE WAS INSPECTED. RECREATION OF EVENT ON (B) (6) 2008 AT DEVICE MFG FACILITY. UPON INVESTIGATION, IT WAS NOTED THAT THE SET SCREW ON THE ARM SWIVEL HOUSING, WHICH IS LOCTITED, THE LOCTITE DOES NOT ADHERE PROPERLY ONCE ARM ASSEMBLY IS INSTALLED WITH GREASE. DEVICE WAS REPLACED WITH NEW ARM ASSEMBLY. CORRECTIVE ACTION WAS GENERATED FROM INVESTIGATION. DEVICE WAS RETURNED TO PT (B) (6) IN (B) (6) 2008, WHERE HE WAS ABLE TO CONTINUE USE OF THE CHAIR. ISSUE WAS RESOLVED.

Description of Event or Problem · 1

PT'S MOM CALLED IN AND SAID THAT HER SON, PT: (B) (6), FELL OUT OF HIS STANDING WHEELCHAIR (MODEL: HPS-2, (B) (4)), AND HIT HIS HEAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HALF POWER SUPERSTAND STANDUP WHEELCHAIR STANDUP WHEELCHAIR IPL THE STANDING COMPANY HPS-2 NA

Patients

Seq Age Sex Outcome Treatment
1 43 YR Other HPS-2, 2 YRS OF USE| HALF POWER SUPERSTAND STANDUP WHEELCHAIR