FDA Adverse Event Malfunction Summary report: N

INFANT INTESNIVE CARE SYSTEM

MDR report key: 1712798 · Received June 4, 2010

Report

Report Number
2510954-2010-00005
Event Type
Malfunction
Date Received
June 4, 2010
Date of Event
March 1, 2010
Report Date
June 3, 2010
Manufacturer
DRAEGER MEDICAL SYSTEMS, INC. (PNC)
Product Code
FMT
PMA / PMN Number
KB75270
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DRAEGER COMPLETED AN INVESTIGATION AND CONFIRMED THIS REPORT. THE HEAT STRESS DAMGE WAS CAUSED BY HIGH CONTACT RESISTANCE AT PIN 1 OF THE J5 CONNECTOR ON THE PRINTED CIRCUIT BOARD. THE THERMAL STRESS OCCURRED WITH THE AGING OF THE CONNECTOR. THIS FAILURE MODE RESULTED IN A MEDICAL DEVICE BULLETIN. THE DEVICE MATERIAL MEET THE APPLICABLE REQUIREMENTS OF UL 94V CLASSIFICATIONS AND ARE SELF EXTINGUISHING. THE POWER MODULE IN WHICH THE FAILURE OCCURRED IS SELF CONTAINED WITHIN THE WARMER MODULE. THE (B)(4) IS AN OPEN AIR DEVICE. DRAEGER WILL CONTINUE TO MONITOR FOR SIMILAR REPORTS OF THIS CONDITION.

Description of Event or Problem · 1

A USER FACILITY REPORT WAS REC'D BY DRAEGER ON (B)(6) 2010 INDICATING A UNIT HAD AN INFANT IN A BASSINET, THE UNIT STARTED SMOKING IN THE AREA OF THE WARMER POWER MODULE. THE UNIT WAS REMOVED FROM SERVICE. NO PT INJURY WAS REPORTED. REFERENCE MFG COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFANT INTESNIVE CARE SYSTEM INFANT WARMER FMT DRAEGER MEDICAL SYSTEMS, INC. (PNC) IICS90 7830 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK