670G INSULIN PUMP MMT-1780KL
Report
- Report Number
- 2032227-2023-217668
- Event Type
- Malfunction
- Date Received
- June 14, 2023
- Date of Event
- May 20, 2023
- Report Date
- September 26, 2023
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000365882
- PMA / PMN Number
- P160017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
INSERTED A TEST P-CAP INTO THE RETAINER RING, AND IT LOCKED IN PLACE PROPERLY. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: CRACKED BATTERY TUBE THREADS, CRACKED CASE ON THE BATTERY TUBE SIDE THAT STARTS AT THE LEFT BELT CLIP RAIL, ANOTHER CRACK IN THE CASE THAT RUNS HORIZONTALLY ACROSS THE BATTERY TUBE ABOUT WHERE THE BOTTOM OF THE BATTERY COMPARTMENT IS LOCATED, MISSING DISPLAY WINDOW COVER, CRACKED MIDDLE (ENTER) KEYPAD BUTTON, SCRATCHED KEYPAD OVERLAY, KEYPAD OVERLAY TEXTURE DAMAGE, SCRATCHED CASE. THE PUMP WAS RECEIVED WITH A BATTERY CAP. THE CONTACT WAS PROPERLY ATTACHED, AND THE WELD SPOTS HOLDING IT IN PLACE WERE STILL INTACT. AFTER BATTERY INSTALLATION, A BLANK DISPLAY WAS NOTED. THE CASE WAS CUT OPEN. AFTER VISUAL INSPECTION, DID NOT NOTE ANY SIGNS OF PREVIOUS MOISTURE PRESENCE OR CORROSION INSIDE THE BATTERY COMPARTMENT OR ON ANY OF THE BOARDS, ASSEMBLIES, AND THE MOTOR INSIDE THE PUMP. THERE IS A CRACK, HOWEVER, AT THE BOTTOM OF THE PLASTIC WHICH HOUSES THE BATTERY COMPARTMENT. THE COPPER SLEEVE WITHIN THE BATTERY COMPARTMENT GOT PUSHED DOWNWARDS, AND AS A RESULT, THE BATTERY CAP TERMINAL DID NOT MAKE CONTACT WITH THE BATTERY SLEEVE. THE BATTERY COMPARTMENT WAS THEN PUSHED UP BACK INTO POSITION. THE PUMP THEN BOOTED UP NORMALLY. THE SOFTWARE VERSION WAS THEN ABLE TO BE OBTAINED. THE PUMP PASSED THE DISPLACEMENT, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, AND SELF TESTS. THUS SOFTWARE WAS UTILIZED AND UPLOADED TRACE/HISTORY FILES PROPERLY. IN SUMMARY, THE CUSTOMER¿S REPORT OF A BLANK DISPLAY AND A CRACK IN THE CASE BOTH WERE CONFIRMED DURING TESTING. THE BLANK DISPLAY IS DUE TO THE LACK OF CONTACT BETWEEN THE BATTERY CAP CONTACTS AND THE BATTERY SLEEVE. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS NOT TURNED ON AND ALSO HAD A CRACK ON THE BATTERY COMPARTMENT. TROUBLESHOOTING WAS PERFORMED HOWEVER THE DISPLAY COULD NOT TURN BACK ON. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE USING THE DEVICE AND WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1972876 | 670G INSULIN PUMP MMT-1780KL | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1780KL | HG4E7AP | 000000763000365882 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 24 YR | Female |