FDA Adverse Event Malfunction Summary report: N

AUTOLOG IQ

MDR report key: 17117326 · Received June 13, 2023

Report

Report Number
2184009-2023-00502
Event Type
Malfunction
Date Received
June 13, 2023
Date of Event
March 2, 2021
Report Date
June 13, 2023
Manufacturer
PERFUSION SYSTEMS
Product Code
CAC
UDI-DI
00643169982864
PMA / PMN Number
K181954
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION SUMMARY: THE REPORTED PERFORMANCE ISSUE WAS VERIFIED DURING SERVICE. MEDTRONIC SERVICE REPORTED THAT DURING TESTING, IT WAS FOUND THAT THE OPTICS HAD FAILED. THE HLT FAILED AT THE POINT WHERE LED 11 DID NOT REACT PROPERLY. THE ISSUE WAS RESOLVED BY REPLACING THE EMITTER, DETECTOR AND CABLES. PREVENTIVE MAINTENANCE WAS PERFORMED PER SPECIFICATIONS. CONCLUSION: REPORTED EVENT WAS CONFIRMED. DUE TO THE AVAILABLE INFORMATION AND THE AGE OF THE INSTRUMENT THE MOST LIKELY CAUSE IS NO RMAL COMPONENT WEAR OUT. ALL ASSOCIATED RISKS ARE LOW AND NO DEVICE RELATED PATIENT/CLINICAL SAFETY ISSUES WERE REPORTED. TRENDS FOR ISSUES WITH THIS PRODUCT ARE REVIEWED AT QUARTERLY QUALITY MEETINGS. THIS REGULATORY REPORT IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW AND REMEDIATION PER D00953163 AS PART OF A CAPA ACTION. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION THAT DURING USE THIS AUTOLOG INSTRUMENT IT DUMPED THE ENTIRE CONTENTS INTO THE WASTE BAG (PER PERFUSIONIST - 'EMPTIED CARDIOTOMY STRAIGHT TO WASTE BAG'). USE OF INSTRUMENT WAS ASKED BUT IS UNKNOWN AND THERE WAS NO REPORTED ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED FROM THE SERVICE TECHNICIAN WHICH STATED THAT THERE WAS NO ISSUE WITH THE WASH KIT AND THAT THIS ISSUE WAS WITH THE INSTRUMENT'S OPTICS. ADDITIONAL INFORMATION: IT WAS ASKED BUT IS UNKNOWN IF A TRANSFUSION WAS REQUIRED DUE TO THIS EVENT. HOWEVER, THERE WAS NO ADVERSE PATIENT IMPACT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
171468 AUTOLOG IQ APPARATUS, AUTOTRANSFUSION CAC PERFUSION SYSTEMS ATLGIQ 00643169982864

Patients

Seq Age Sex Outcome Treatment
1 Unknown