XKNIFE RT LINUX WORKSTATION
Report
- Report Number
- 1222895-2010-00004
- Event Type
- Malfunction
- Date Received
- May 20, 2010
- Report Date
- May 20, 2010
- Manufacturer
- INTEGRA RADIONICS
- Product Code
- KPQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HU
- Reporter Occupation
- OTHER
Narratives
THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS BEEN RECEIVED FOR EVAL. AN INVESTIGATION HAS BEEN INITIATED BASED UPON THE REPORTED INFO. INTEGRA ENGINEERS COMPLETED A PRELIMINARY EVAL OF THIS DEVICE AND THE FOLLOWING WAS DETERMINED; THE PRELIMINARY INVESTIGATION DETERMINED THAT THE "DIAMETER" PARAMETER FOR THE 30.0 MM CONE IN THE SITES' BEAM DATA FILE HAD BEEN INCORRECTLY SET TO 37.5 MM. OTHER PARAMETERS FOR THE 30.0 MM CONE WERE FOUND TO BE CORRECT. ADDITIONAL INFO HAS BEEN REQUESTED TO COMPLETE THE INVESTIGATION.
XKNIFE RT LINUX WORKSTATION; THE OUTLINE OF THE 3 CM CONE IN THE SOFTWARE APPEARED BIGGER THAN THE 3.5 CM CONE. THE INCIDENT WAS DESCRIBED AS FOLLOWS BY THE REPORTING PHYSICIAN; "DURING PLANNING IN THE ARC MODE IT WAS NOTED THAT THE (CIRCULAR) COLLIMATOR OUTLINE OF 30 MM DIAMETER FIELD IS BIGGER THAN THE 32.5 MM FIELD. I CHECKED THE DOSE DISTRIBUTION FOR BOTH FIELDS AND THEY SEEM OK, BUT IT LOOKS STRANGE AND A LITTLE BIT ANNOYING FOR OUR NEUROSURGEON. FOR THE 30 MM FIELD THIS COLLIMATOR OUTLINE IS EQUAL WITH ABOUT THE 50% ISODOSE LINE AND ARE EQUAL WITH ABOUT THE 80% FOR THE OTHERS. INTERESTINGLY WE HAVE THIS PROBLEM ONLY FOR THIS FIELD DIAMETER." THE PHYSICIAN WAS ASKED IF ANY PTS WERE TREATED IF ANY INJURY OCCURRED. THE PHYSICIAN STATED THAT 4 MALES AND 1 FEMALE WERE TREATED USING THIS CONE BUT NONE OF THEM WAS INJURED BECAUSE "WE CHECKED EACH CASE CAREFULLY THE ISODOSE CURVES AND CHOSE THE DIAMETER ACCORDING TO ISODOSES AND NOT TO THE COLLIMATOR OUTLINE." ADDITIONAL CLINICAL INFO HAS BEEN REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | XKNIFE RT LINUX WORKSTATION | NA | KPQ | INTEGRA RADIONICS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |