FDA Adverse Event Injury Summary report: N

XIENCE PRIME EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 17098593 · Received June 9, 2023

Report

Report Number
2024168-2023-06100
Event Type
Injury
Date Received
June 9, 2023
Date of Event
May 17, 2023
Report Date
January 10, 2024
Manufacturer
ABBOTT VASCULAR
Product Code
NIQ
PMA / PMN Number
P110019
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING.D4: LOT NUMBER UPDATED FROM 2081742 TO 2090741.

Additional Manufacturer Narrative · 0

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. A CONCLUSIVE CAUSE FOR THE REPORTED PATIENT EFFECT, AND THE RELATIONSHIP TO THE PRODUCT, IF ANY, CANNOT BE DETERMINED; HOWEVER, THE SUBSEQUENT TREATMENT APPEARS TO BE RELATED TO THE OPERATIONAL CONTEXT OF THE PROCEDURE. THE REPORTED PATIENT EFFECT OF DISSECTION IS LISTED IN THE XIENCE PRIME, XIENCE PRIME SMALL VESSEL (SV), AND XIENCE PRIME LONG LENGTH (LL), EVEROLIMUS ELUTING CORONARY STENT SYSTEMS INSTRUCTIONS FOR USE AS A KNOWN PATIENT EFFECT OF CORONARY STENTING PROCEDURES. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A MODERATELY CALCIFIED, MILDLY TORTUOUS, 80% STENOSED, DE NOVO LESION IN THE LEFT MAIN CORONARY ARTERY (LMCA). A 3.5X12MM XIENCE PRIME DRUG ELUTING STENT (DES) WAS IMPLANTED AT THE TARGET LESION AND POST-DILATATION WAS COMPLETED WITH AN NC BALLOON, AFTER WHICH A DISSECTION WAS NOTED. A 3X08MM XIENCE PRIME DES WAS USED TO TREAT THE DISSECTION AND SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
494942 XIENCE PRIME EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT DELIVERY SYSTEM NIQ ABBOTT VASCULAR 2090741

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention