FDA Adverse Event Malfunction Summary report: N

PKG, CROSSFIRE 2 CONSOLE

MDR report key: 17092710 · Received June 8, 2023

Report

Report Number
0002936485-2023-00571
Event Type
Malfunction
Date Received
June 8, 2023
Date of Event
May 12, 2023
Report Date
February 16, 2024
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
GEI
UDI-DI
07613327058109
PMA / PMN Number
K071859
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 0

ALLEGED FAILURE: EIB, CHASE, REP, CROSSFIRE CONSOLE STARTED SMOKING WHILE ABLATOR WAS BEING USED IN A SHOULDER ARTHROSCOPY CASE., TEMP. *DELAY: WE HAD TO SWITCH TOWERS AND CHANGE CONSOLES. THE FAILURE(S) IDENTIFIED IN THE INVESTIGATION IS CONSISTENT WITH THE COMPLAINT RECORD. THE ROOT CAUSE IS THE RF BOARD. THE PRODUCT WAS RETURNED FOR INVESTIGATION AND THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
697380 PKG, CROSSFIRE 2 CONSOLE ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES GEI STRYKER ENDOSCOPY-SAN JOSE 0475100000 07613327058109

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown