FDA Adverse Event Malfunction Summary report: N

VORTEK HYDRO-COATED DOUBLE LOOP URETERAL STENT

MDR report key: 17091196 · Received June 8, 2023

Report

Report Number
9610711-2023-00119
Event Type
Malfunction
Date Received
June 8, 2023
Report Date
August 31, 2023
Manufacturer
COLOPLAST A/S
Product Code
FAD
PMA / PMN Number
K170362
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

WE HAVE BEEN INFORMED ABOUT A DEFECT PRODUCT: JJ IN CH. 8 IN THE PACKAGING ACCORDING TO THE COMPLAINT DESCRIPTION, THE SAMPLE ARE AVAILABLE AND THE LOT NUMBER. AFTER RECEIVING THIS COMPLAINT, WE SEARCHED FOR OTHER COMPLAINTS WITH THE SAME LOT NUMBER 9070782 WE FOUND FOUR OTHER COMPLAINTS CHECKING THE QUALITY DATABASE REVEALED NO ANOMALY IN RELATION WITH THE DESCRIBE DEFECT. ON JULY WE RECEIVED SEALED SAMPLE. AFTER OBSERVATION WE NOTE THAT JJ ITEM NUMBER YCV284 CH8 WERE PACKAGED INSTEAD OF JJ ITEM NUMBER YCH17490 CH7 WERE PACKAGED. PRODUCTION UNIT FOR THIS PRODUCT WAS INFORMED ABOUT THIS DEFECT. AN AWARENESS WAS ALREADY DONE ABOUT THIS TOPIC ON MAY. SIMILAR CASE STUDY WAS DONE BASED ON SAME ITEM NUMBER, SAME DEFECT ( PACKAGING/LABELLING ERROR), OVER LAST FOUR YEAR, AND 5 OTHERS COMPLAINTS ( SAME ISSUE BUT DUPLICATE FOR EMDR).

Additional Manufacturer Narrative · 0

D4 LOT NUMBER: 9070782 WITHOUT THE BENEFIT OF EXAMINATION AND TESTING, COLOPLAST IS PRECLUDED FROM COMMENTING ON THE CONDITION OF THE DEVICE OR THE CAUSE OF THE OCCURRENCE. SHOULD ADDITIONAL INFORMATION PROMPT US TO ALTER OR SUPPLEMENT ANY INFORMATION OR CONCLUSIONS CONTAINED IN THIS REPORT, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

ACCORDING TO THE AVAILABLE INFORMATION, THE ITEM NUMBER WAS A 7 CH/FR DEVICE; HOWEVER, THE STENT IN THE PACKAGE WAS 8 CH/FR. THE PUSHER WAS 7 CH/FR, SO THE CONNECTION DID NOT FIT INTRAOPERATIVELY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
270706 VORTEK HYDRO-COATED DOUBLE LOOP URETERAL STENT STENT, URETERAL FAD COLOPLAST A/S BCFG741002

Patients

Seq Age Sex Outcome Treatment
1 Unknown