FDA Adverse Event Malfunction Summary report: N

TOTALCARE-TREAT SURFACE&SCALE TOTAL CARE FRAME

MDR report key: 1708596 · Received August 20, 2009

Report

Report Number
1824206-2009-04770
Event Type
Malfunction
Date Received
August 20, 2009
Date of Event
October 18, 2006
Report Date
October 18, 2006
Manufacturer
HILL-ROM RITTER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN USED SIDECOM TESTER AND FOUND SIDECOM PC BOARD BAD. TECH REPLACED BOARD AND PERFORMED OPERATION CHECK. ALL PASS. PURSUANT TO OUR RESPONSE TO WARNING LETTER 2008-DT-04, HILL-ROM IS CONTINUING ITS RETROSPECTIVE REVIEW OF EVENTS FOR REPORTABILITY. THIS EFFORT WILL CONTINUE UNTIL WE ARE CURRENT.

Description of Event or Problem · 1

BED NOT SENDING NURSE CALL OUT TO STATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE-TREAT SURFACE&SCALE TOTAL CARE FRAME AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM RITTER P1900AF1619

Patients

Seq Age Sex Outcome Treatment
1