FDA Adverse Event
Injury
Summary report: N
STONE CRUSHING FORCEPS, SGL. ACTION JAWS
MDR report key: 17079695
·
Received June 7, 2023
Report
- Report Number
- 9610617-2023-00912
- Event Type
- Injury
- Date Received
- June 7, 2023
- Date of Event
- September 27, 2017
- Report Date
- June 7, 2023
- Manufacturer
- KARL STORZ SE & CO. KG
- Product Code
- OCZ
- PMA / PMN Number
- K950434
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
BELATED EVALUATION AND REPORTING OF THIS COMPLAINT WAS DONE DURING RETROSPECTIVE REVIEW AS PART OF CAPA 20-0074 CORRECTIVE ACTION 6. THE MOVABLE JAW IS BROKEN ON BOTH HINGE SIDES. THE FRACTURE SURFACES SHOW THE TYPICAL APPEARANCE OF AN OVERLOAD FRACTURE AND ARE LOCATED AT THE POINT EXPECTED FOR A MECHANICAL OVERLOAD. THE EVENT IS FILED UNDER INTERNAL KARL STORZ COMPLAINT ID (B)(4).
Description of Event or Problem · 0
IT WAS REPORTED THAT THERE WAS EVENT WITH A STONE CRUSHING FORCEPS, SGL. ACTION JAWS. ACCORDING TO THE INFORMATION RECEIVED, DURING MANUAL LITHOTRIPSY FOR BLADDER STONE CASE, MOBILE JAW BROKE. ADDITIONAL PATIENT INFORMATION IS NOT AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 587138 | STONE CRUSHING FORCEPS, SGL. ACTION JAWS | STONE CRUSHING FORCEPS, SGL. ACTION JAWS | OCZ | KARL STORZ SE & CO. KG | 27074B | WT01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |