FDA Adverse Event Injury Summary report: N

STONE CRUSHING FORCEPS, SGL. ACTION JAWS

MDR report key: 17079695 · Received June 7, 2023

Report

Report Number
9610617-2023-00912
Event Type
Injury
Date Received
June 7, 2023
Date of Event
September 27, 2017
Report Date
June 7, 2023
Manufacturer
KARL STORZ SE & CO. KG
Product Code
OCZ
PMA / PMN Number
K950434
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

BELATED EVALUATION AND REPORTING OF THIS COMPLAINT WAS DONE DURING RETROSPECTIVE REVIEW AS PART OF CAPA 20-0074 CORRECTIVE ACTION 6. THE MOVABLE JAW IS BROKEN ON BOTH HINGE SIDES. THE FRACTURE SURFACES SHOW THE TYPICAL APPEARANCE OF AN OVERLOAD FRACTURE AND ARE LOCATED AT THE POINT EXPECTED FOR A MECHANICAL OVERLOAD. THE EVENT IS FILED UNDER INTERNAL KARL STORZ COMPLAINT ID (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS EVENT WITH A STONE CRUSHING FORCEPS, SGL. ACTION JAWS. ACCORDING TO THE INFORMATION RECEIVED, DURING MANUAL LITHOTRIPSY FOR BLADDER STONE CASE, MOBILE JAW BROKE. ADDITIONAL PATIENT INFORMATION IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
587138 STONE CRUSHING FORCEPS, SGL. ACTION JAWS STONE CRUSHING FORCEPS, SGL. ACTION JAWS OCZ KARL STORZ SE & CO. KG 27074B WT01

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other