FDA Adverse Event Malfunction Summary report: N

BREMER CROWN W/ HIFIX SKULL PINS

MDR report key: 1707968 · Received May 28, 2010

Report

Report Number
1526439-2010-00079
Event Type
Malfunction
Date Received
May 28, 2010
Date of Event
May 20, 2010
Manufacturer
DEPUY SPINE INC
Product Code
JEC
PMA / PMN Number
K915800
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CROWN WAS RETURNED WITH VEST CONNECTING BOLT BENT AND BROKEN OFF WITHIN THE CROWN. IT APPEARS THAT THE USER LIKELY THREAD THE BOLT DURING INSERTION. AS REPORTED THE BOLT WAS ADVANCED FURTHER WHEN RESISTANCE WAS FELT CAUSING IT TO BIND. THE ADDITIONAL APPLICATION OF A TYPICAL FORCE RESULTED IN THE BOLT BREAKING. THE THREADED INSERTION HOLES ARE INSPECTED DURING MANUFACTURE TO ENSURE THAT THE THREADS CONFORM TO SPECIFICATION.

Description of Event or Problem · 1

DOCTOR WAS PLACING A LARGE HALO AND AIRFLO VEST. WHEN INSERTING ONE OF THE CRANIAL PINS HE NOTICED THAT IT BECAME DIFFICULT TO THREAD INTO THE CROEN. AS IF IT WAS MIS-THREAD. AFTER TRYING TO CONTINUE TO ADVANCE THE PIN, THE HEAD OF THE BOLT SHEARED OFF. AS A RESULT OF THIS EVENT, THIS WAS A SIGNIFICANT DELAY AND THE ANOTHER CROWN HAD TO BE PLACED ON THE PT THE FOLLOWING DAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BREMER CROWN W/ HIFIX SKULL PINS HALO CROWN JEC DEPUY SPINE INC NA 060871

Patients

Seq Age Sex Outcome Treatment
1 UNK