INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE
Report
- Report Number
- 1037905-2023-00274
- Event Type
- Malfunction
- Date Received
- June 7, 2023
- Date of Event
- May 15, 2023
- Report Date
- June 7, 2023
- Manufacturer
- WILSON-COOK MEDICAL INC
- Product Code
- PKL
- UDI-DI
- 00827002580107
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
PMA/510(K): K21323. INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE REPORT COULD NOT BE CONFIRMED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: A CORRECTIVE ACTION HAS BEEN INITIATED TO REDUCE OCCURRENCES OF THE CLIP HOUSING DETACHING FROM THE CATH ATTACH FOR INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES. THE PRODUCT SAID TO BE INVOLVED IS INCLUDED IN THE SCOPE OF THE CORRECTIVE ACTIONS. THE INSTRUCTIONS FOR USE STATES: "UNCOIL DEVICE. VERIFY SMOOTH HANDLE OPERATION AND CLIP ACTION. OPEN CLIP BY GENTLY MOVING HANDLE SPOOL DISTALLY (AWAY FROM HANDLE THUMB RING). ONCE CLIP IS FULLY OPEN, DO NOT CONTINUE ADVANCING HANDLE SPOOL AS CLIP MAY PREMATURELY DETACH FROM CATHETER. CLOSE CLIP BY MOVING HANDLE SPOOL PROXIMALLY UNTIL CLIP IS FULLY CLOSED. PRECAUTION: DO NOT CONTINUE TO PULL HANDLE SPOOL BEYOND TACTILE RESISTANCE AS THIS MAY PREMATURELY DEPLOY CLIP.". THE INSTRUCTIONS FOR USE STATES: "WITH CLIP CLOSED AND WITHOUT HOLDING HANDLE SPOOL, ADVANCE DEVICE IN SMALL INCREMENTS INTO ACCESSORY CHANNEL OF GASTROSCOPE, DUODENOSCOPE, OR COLONOSCOPE. NOTE: IF DIFFICULT TO ADVANCE DEVICE, RELAX ELEVATOR AND/OR STRAIGHTEN ENDOSCOPE.¿. FAILURE TO FOLLOW THE INSTRUCTIONS ABOVE CAN RESULT IN DAMAGE TO THE DEVICE WHICH MAY LEAD TO PREMATURE CLIP DEPLOYMENT. PRIOR TO DISTRIBUTION, ALL INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MANUFACTURED PRIOR TO IMPLEMENTATION OF THIS CORRECTIVE ACTION. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.
DURING AN UPPER ENDOSCOPIC SUBMUCOSAL DISSECTION, ESD, THE PHYSICIAN USED TWO COOK INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES. IT WAS REPORTED THAT WHEN ADVANCING THROUGH A SIDE VIEWING SCOPE AND EXITING THE SCOPE THE CLIP HOUSING WAS DETACHED FROM THE CATH ATTACH [MOVED IN A TROMBONING MOTION]. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 607337 | INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE | PKL, LIGATOR, HEMORRHOIDAL | PKL | WILSON-COOK MEDICAL INC | G58010 | W4708234 | 00827002580107 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | OLYMPUS SIDE VIEWING SCOPE, UNKNOWN MODEL |