BLUNT TIP SCREW, 4X44MM
Report
- Report Number
- 0009613350-2023-00252
- Event Type
- Injury
- Date Received
- June 7, 2023
- Date of Event
- May 2, 2023
- Report Date
- June 22, 2023
- Manufacturer
- ZIMMER SWITZERLAND MANUFACTURER GMBH
- Product Code
- HSB
- UDI-DI
- 00889024505469
- PMA / PMN Number
- K181827
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
(B)(4). D10: ITEM#:47-2496-160-11 ;LOT#:3127388 ;ITEM NAME: PROXIMAL HUMERUS, RIGHT, 11X160MM. ITEM#:47-2486-042-40 ;LOT#:3091519 ;ITEM NAME: BLUNT TIP SCREW, 4X42MM. ITEM#:47-2486-042-40 ;LOT#:3131295 ;ITEM NAME: BLUNT TIP SCREW, 4X42MM. G2-FOREIGN- JAPAN MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0009613350 -2023 -00246, 0009613350 -2023 -00250, AND 0009613350 -2023 -00251. INVESTIGATION OF THIS INCIDENT IS CURRENTLY ONGOING. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL MEDWATCH 3500A WILL BE SUBMITTED.
(B)(4). THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND/OR CORRECTED INFORMATION. NO PRODUCT WAS RETURNED OR PICTURES PROVIDED; VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. A REVIEW OF THE DEVICE MANUFACTURING RECORDS CONFIRMED NO ABNORMALITIES OR DEVIATIONS. DEVICES ARE USED FOR TREATMENT. MEDICAL RECORDS WERE NOT PROVIDED. IT CAN BE ASSUMED THAT MULTIPLE CONTRIBUTING FACTORS, RELATED TO EITHER PATIENT CONDITION AND BEHAVIOR OR IMPLANTATION PROCEDURE, CONTRIBUTED TO THE MIGRATION OF THE SCREWS. IF AND TO WHAT EXTENT ANY OF THESE ASPECTS MAY HAVE INFLUENCED THE MIGRATION OF THE SCREW REMAINS UNKNOWN. AN ADDITIONAL DEEPER INVESTIGATION WAS PERFORMED WHICH IDENTIFIED THE DESIGN LIMITATION OF THE CORELOCK MECHANISM OF THE AFFIXUS NAIL AS A POTENTIAL CONTRIBUTING FACTOR. AS PART OF THE DEEPER INVESTIGATION, A SCIENTIFIC LITERATURE SEARCH WAS CONDUCTED AND A SYSTEMATIC REVIEW OF THE OUTCOME OF INTRAMEDULLARY NAILING FOR ACUTE PROXIMAL HUMERUS FRACTURE SHOWED THAT SCREW MIGRATION AND PERFORATION INTO THE JOINT IS THE SECOND MOST COMMON COMPLICATION FOLLOWING SECONDARY LOSS OF REDUCTION. IN ADDITION, SCIENTIFIC LITERATURE OF COMPETITOR AND PREDICATE DEVICES WERE ASSESSED. THIS REVIEW HAS SHOWN THAT THE AFFIXUS® NATURAL NAIL® PROXIMAL HUMERAL SYSTEM PERFORMS BETTER AND/OR IN EQUAL RANGE IN COMPARISON WITH LEGALLY MARKETED SIMILAR DEVICES. IN CONCLUSION, AS THE CAUSE MAY BE MULTIFACTORIAL AN EXACT ROOT CAUSE COULD NOT BE IDENTIFIED FOR THE MIGRATION OF THE SCREW. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THAT DURING THE INITIAL SURGERY, THE PATIENT WAS IMPLANTED WITH ANN NAIL. 6 WEEKS AFTER THE SURGERY ONE OF THE PROXIMAL SCREWS WAS BACKED OUT FROM THE PROPER POSITION AND THE PATIENT EXPERIENCED PAIN. SUBSEQUENTLY, THE PATIENT UNDERWENT REVISION SURGERY TO REMOVE THE SCREW. DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 482808 | BLUNT TIP SCREW, 4X44MM | ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES | HSB | ZIMMER SWITZERLAND MANUFACTURER GMBH | 47-2486-044-40 | 3134549 | 00889024505469 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Prefer Not To Disclose | Hospitalization| R | SEE H10 NARRATIVE. |