DAVINCI XI
Report
- Report Number
- 2955842-2023-15844
- Event Type
- Malfunction
- Date Received
- June 6, 2023
- Date of Event
- June 22, 2021
- Report Date
- June 22, 2021
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874110744
- PMA / PMN Number
- K131861
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- 003
Narratives
AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED COMPLAINT. THE FSE REPLACED THE HIGH-RESOLUTION STEREO VIEWER (HRSV) MONITOR TO RESOLVE THE ISSUE. NO ERRORS WERE FOUND IN THE LOG FILES. THE FSE PERFORMED ELECTRICAL SAFETY TESTS AND VERIFIED THE SYSTEM AFTER THE REPAIR. THE SYSTEM WAS RE-TESTED AND VERIFIED AS READY FOR USE. INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE HIGH-RESOLUTION STEREO VIEWER (HRSV) MONITOR INVOLVED WITH THIS COMPLAINT AND COMPLETED THE DEVICE EVALUATION. FAILURE ANALYSIS (FA) INVESTIGATION CONFIRMED THE CUSTOMER REPORTED COMPLAINT. FAILURE ANALYSIS INVESTIGATION FOUND NO IMAGE ON THE MONITOR DUE TO THE MAIN BOARD WAS DEFECTIVE.
IT WAS REPORTED THAT DURING AN UNSPECIFIED DA VINCI-ASSISTED SURGICAL PROCEDURE, THE SURGEON SIDE CONSOLE (SSC) HIGH-RESOLUTION STEREO VIEWER (HRSV) RIGHT EYE HAD NO VISION. THE SURGEON MADE THE DECISION TO COMPLETE THE PROCEDURE USING A SECOND SSC WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE FIELD SERVICE ENGINEER (FSE) AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE SYSTEM IS ALWAYS CONFIGURED AS A DUAL CONSOLE CONFIGURATION. THE SYSTEM INITIALLY POWERED ON WITHOUT ERRORS AND FUNCTIONALITY WAS CHECKED. THE CUSTOMER EXPERIENCED THE RIGHT EYE VISION LOSS MID PROCEDURE. THE CUSTOMER CONTACTED A FSE AT THE TIME OF THE REPORTED PROBLEM; HOWEVER, FULL TROUBLESHOOTING WAS NOT COMPLETED BECAUSE THE SURGEON PREFERRED TO SWITCH TO THE SECOND CONSOLE RATHER THAN CAUSING DELAYS. THE FSE COMPLETED TROUBLESHOOTING AFTER THE PROCEDURE AND FOUND THE SSC HRSV REQUIRED REPLACEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 745119 | DAVINCI XI | SURGEON SIDE CONSOLE, SMART PEDALS | NAY | INTUITIVE SURGICAL, INC | 380677-14 | N/A | 00886874110744 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES |