FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 17067636 · Received June 6, 2023

Report

Report Number
2955842-2023-15844
Event Type
Malfunction
Date Received
June 6, 2023
Date of Event
June 22, 2021
Report Date
June 22, 2021
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110744
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED COMPLAINT. THE FSE REPLACED THE HIGH-RESOLUTION STEREO VIEWER (HRSV) MONITOR TO RESOLVE THE ISSUE. NO ERRORS WERE FOUND IN THE LOG FILES. THE FSE PERFORMED ELECTRICAL SAFETY TESTS AND VERIFIED THE SYSTEM AFTER THE REPAIR. THE SYSTEM WAS RE-TESTED AND VERIFIED AS READY FOR USE. INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE HIGH-RESOLUTION STEREO VIEWER (HRSV) MONITOR INVOLVED WITH THIS COMPLAINT AND COMPLETED THE DEVICE EVALUATION. FAILURE ANALYSIS (FA) INVESTIGATION CONFIRMED THE CUSTOMER REPORTED COMPLAINT. FAILURE ANALYSIS INVESTIGATION FOUND NO IMAGE ON THE MONITOR DUE TO THE MAIN BOARD WAS DEFECTIVE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING AN UNSPECIFIED DA VINCI-ASSISTED SURGICAL PROCEDURE, THE SURGEON SIDE CONSOLE (SSC) HIGH-RESOLUTION STEREO VIEWER (HRSV) RIGHT EYE HAD NO VISION. THE SURGEON MADE THE DECISION TO COMPLETE THE PROCEDURE USING A SECOND SSC WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE FIELD SERVICE ENGINEER (FSE) AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE SYSTEM IS ALWAYS CONFIGURED AS A DUAL CONSOLE CONFIGURATION. THE SYSTEM INITIALLY POWERED ON WITHOUT ERRORS AND FUNCTIONALITY WAS CHECKED. THE CUSTOMER EXPERIENCED THE RIGHT EYE VISION LOSS MID PROCEDURE. THE CUSTOMER CONTACTED A FSE AT THE TIME OF THE REPORTED PROBLEM; HOWEVER, FULL TROUBLESHOOTING WAS NOT COMPLETED BECAUSE THE SURGEON PREFERRED TO SWITCH TO THE SECOND CONSOLE RATHER THAN CAUSING DELAYS. THE FSE COMPLETED TROUBLESHOOTING AFTER THE PROCEDURE AND FOUND THE SSC HRSV REQUIRED REPLACEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
745119 DAVINCI XI SURGEON SIDE CONSOLE, SMART PEDALS NAY INTUITIVE SURGICAL, INC 380677-14 N/A 00886874110744

Patients

Seq Age Sex Outcome Treatment
1 Unknown DA VINCI INSTRUMENTS AND ACCESSORIES